6 March 2013 A simulation based image reconstruction strategy with predictable image quality in limited-angle x-ray tomography
Author Affiliations +
Proceedings Volume 8668, Medical Imaging 2013: Physics of Medical Imaging; 86685P (2013) https://doi.org/10.1117/12.2007976
Event: SPIE Medical Imaging, 2013, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
This paper presents a Pre-computed BackProjection (BP) based Penalized-likelihood (PPL) method for limited­ angle X-ray tomography based on the theory of resolution properties of regularized image reconstruction. Pre­ computed BP based penalty is a simplified version of the modified quadratic penalty proposed in the literature. 1 By inserting a BP equivalent estimation into a quadratic penalty, the data-related terms in the impulse response and noise reconstructed by PPL are absorbed, such that the effects of smoothing parameter of the penalty can be evaluated in advance through the simulated data. A simulation based two-step procedure is proposed to apply PPL method in real applications. It reconstructs images with predictable resolution properties by choosing a corresponding smoothing parameter. The effectiveness and robustness of the two-step strategy is validated through simulation based experiments.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Shiyu Xu and Ying Chen "A simulation based image reconstruction strategy with predictable image quality in limited-angle x-ray tomography", Proc. SPIE 8668, Medical Imaging 2013: Physics of Medical Imaging, 86685P (6 March 2013); https://doi.org/10.1117/12.2007976
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Tomography

Image restoration

X-ray imaging

Modulation transfer functions

X-ray detectors

Sensors

RELATED CONTENT

Brute force absorption contrast microtomography
Proceedings of SPIE (September 12 2014)
Large-area mercuric iodide x-ray imager
Proceedings of SPIE (May 03 2002)
Low-exposure scanning-beam x-ray fluoroscopy system
Proceedings of SPIE (April 11 1996)
X-Ray tomography with micrometer spatial resolution
Proceedings of SPIE (October 24 1997)

Back to Top