30 August 2013 Calibration and correction of the CCD spectral response nonuniformity for Fourier transform imaging spectrometer
Dalian Shi, Xuebin Liu, Shuang Wang, Juanjuan Jing, Desheng Wen, Qinglan Bai
Author Affiliations +
Proceedings Volume 8910, International Symposium on Photoelectronic Detection and Imaging 2013: Imaging Spectrometer Technologies and Applications; 89101C (2013) https://doi.org/10.1117/12.2034271
Event: ISPDI 2013 - Fifth International Symposium on Photoelectronic Detection and Imaging, 2013, Beijing, China
Abstract
Due to the manufacturing technique, some kinds of CCD, such as the back illuminated CCD, have the problem of spectral response nonuniformity. The near infrared light passing through the substrate and gates and is reflected back into the substrate for a second pass resulting in increased response. For the Fourier transform imaging spectrometer, it adds stripe pattern error to the interferogram and distorts the reconstructed spectrum. The nonuniform response is wavelength dependent due to changes in reflectivity of metal and the cavity formed by silicon and metal with transparent dielectric, so it adds difficulty to the correction of the error of the reconstructed spectrum.

In order to reduce the error of the reconstructed spectrum, in this paper, a calibration method and a correction method to correct the error caused by the CCD spectral response nonuniformity was developed, basing on analysis of the property of the CCD spectral response nonuniformity. Firstly, a calibrated monochromater was used to measure the CCD spectral response nonuniformity and the property and affect of the CCD spectral response nonuniformity were analyzed. Method to correct the error of the reconstructed spectrum caused by the stripe pattern error was developed. Secondly, to calibrate the CCD spectral response nonuniformity, the spectral response coefficient and the spatial response nonuniformity coefficient was measured and computed. Finally, we took data with a Fourier transform imaging spectrometer, and got the correction results of the reconstructed spectrums. The results showed that the distortion of recovered spectrum was evidently reduced and the effect of the calibration and correction method was proved.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Dalian Shi, Xuebin Liu, Shuang Wang, Juanjuan Jing, Desheng Wen, and Qinglan Bai "Calibration and correction of the CCD spectral response nonuniformity for Fourier transform imaging spectrometer", Proc. SPIE 8910, International Symposium on Photoelectronic Detection and Imaging 2013: Imaging Spectrometer Technologies and Applications, 89101C (30 August 2013); https://doi.org/10.1117/12.2034271
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Charge-coupled devices

Calibration

Spectroscopy

Fourier transforms

Error analysis

Spectral calibration

Distortion

Back to Top