1 September 2008 Double-layer estimation of intra- and extracerebral hemoglobin concentration with a time-resolved system
Louis Gagnon, Claudine Gauthier, Richard D. Hoge, Frederic Lesage, Juliette J. Selb, David A. Boas
Author Affiliations +
Abstract
We present in vivo measurements of baseline physiology from five subjects with a four-wavelength (690, 750, 800, and 850 nm) time-resolved optical system. The measurements were taken at four distances: 10, 15, 25, and 30 mm. All distances were fit simultaneously with a two-layered analytical model for the absorption and reduced scattering coefficient of both layers. The thickness of the first layer, comprising the skin, scalp, and cerebrospinal fluid, was obtained from anatomical magnetic resonance images. The fitting procedure was first tested with simulations before being applied to in vivo measurements and verified that this procedure permits accurate characterization of the hemoglobin concentrations in the extra- and intracerebral tissues. Baseline oxyhemoglobin, deoxyhemoglobin, and total hemoglobin concentrations and oxygen saturation were recovered from in vivo measurements and compared to the literature. We observed a noticeable intersubject variability of the hemoglobin concentrations, but constant values for the cerebral hemoglobin oxygen saturation.

1.

Introduction

Imaging with near-infrared diffuse optical imaging (DOI) in the neurosciences has seen increased interest over the past 15years .1, 2, 3, 4 In the commonly used continuous-wave (CW) version,5, 6 DOI can only measure relative changes in oxy- (HbO2) and deoxyhemoglobin (HbR) concentrations. On the contrary, frequency domain (FD)7, 8, 9, 10 and time domain (TD) 11, 12, 13, 14, 15, 16, 17, 18, 19 technologies enable absolute measurements of the medium’s optical properties.20, 21, 22 This is particularly useful to calibrate brain activation and quantify the underlying hemodynamic processes within the brain. For example, multimodal studies (e.g., optical-MRI fusion23) need quantitative information to estimate the cerebral metabolic rate of oxygen (CMRO2) . In addition, the blood-oxygen level dependant (BOLD) signal depends both on cerebral blood flow (CBF) and CMRO2 . However, this relation is not straightforward, and a calibration constant must be estimated. The change in CBF can be measured separately by arterial spin labeling fMRI so the only two unknowns are the calibration constant and CMRO2 . To estimate CMRO2 , one must measure the BOLD signal at two different CBF values but without altering the CMRO2 . This must be done during hypercapnic periods using two different levels of CO2 pressure.24 However, TD measurements could provide an alternative to this procedure.

Brain optical properties have been measured previously in vitro,25 but physiological factors make those measurements distinct from the in vivo situation. Among those distinctions is the swelling of the mitochondria, which results in structural changes, and the small fluctuations of the brain temperature, which change the hemoglobin solubility in the blood. Moreover, previous studies performing in vivo measurements reported noticable intersubject variability.8, 26 These results preclude the generalization of single-subject measured values to other subjects and confirm the importance of obtaining these values individually for quantitative imaging.

In previous work, baseline optical measurements with a TD system using multidistances and a homogeneous model has been used to fit the data and estimate the optical parameters.26, 27 A major drawback is that this homogeneous model does not distinguish between hemoglobin concentrations in the scalp and those of the cerebral tissues, because it does not account for the layered structure of the head. The superficial layer, the skin and skull, limit the accuracy of DOI, since light is also absorbed and scattered in these regions, which are not part of the cerebral cortex. Moreover, the skin layer is subject to a physiology that may cause interference in the process of recovering cerebral activity. Several methods based on multi-distance measurements have been developed to overcome this problem.28, 29, 30, 31 In a multimodal study combining position emission tomography and TD optical imaging, it was shown that the contributions from these superficial layers are reduced significantly, even using a homogeneous model, when source-detector distances are increased beyond 4cm .32 Such measurements require a large signal-to-noise ratio, which was not available with the system used in our experiment.

Separately, analytical models have been developed by solving the diffusion equation and its boundary conditions for a two-layered medium.22, 33, 34, 35 These models, validated with Monte Carlo simulations, showed adequate efficiencies at recovering the parameters if the thickness of the first layer of the model was known a priori. In vivo measurements with a two-layered model also have been reported using a FD system,8 and as expected, clear distinctions between scalp and cerebral tissues properties were made. The goal here is to provide further confirmation of the above FD results (Choi) with an independent TD technique.

In this work, we report intra- and extracerebral hemoglobin concentrations recovered on individual subjects with a time-resolved system using a two-layered analytical model for the first time. The measurements were taken with four wavelengths (690, 750, 800, and 850nm ) at four distances: 10, 15, 25, and 30mm . All wavelengths were fit simultaneously with a two-layered analytical model for the absorption and reduced scattering coefficient of both layers. Concentrations were then computed with the recovered absorption coefficients. We observed a large variability between subjects. Results were compared to the literature, and differences between time and frequency measurements for the oxygen saturation in the skin and skull layer were observed.

2.

Theory

Light propagation in a turbid medium is described by the radiative transport equation. This equation can be further approximated by the diffusion equation when the medium is highly scattering. Analytical solutions have been developed in the literature for simple geometries and homogeneous media.36 Analytical 35, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46 and numerical33, 47 solutions also have been developed for multilayered media. Only the solutions used in our fitting procedure are described in this section.

2.1.

Homogenous Medium

As given in Kienle, 33 the radiance at the surface of a homogeneous medium at a distance ρ from the photon source following a time t after the source pulse is given by

Eq. 1

R(ρ,t)=0.118Φ(ρ,z=0,t)+0.306Φzz=0,
where Φ is the solution of the diffusion equation given by Patterson 36 with an extrapolated boundary condition,48 and Φ is expressed as

Eq. 2

Φ(ρ,z,t)=Ac(4πDct)32exp(μact){exp[z24Dct]exp[z*24Dct]},
where c=0.214mms1 is the speed of light in tissue, and A is an amplitude factor accounting for the source magnitude. The diffusion coefficient, D , is defined by

Eq. 3

D=13[μa+μs]1,
where μa is the absorption coefficient, and μs is the effective scattering coefficient. Also, z2=z02+ρ2 is the distance between the isotropic photon source and the detector, and z*2=(z0+2zb)2+ρ2 is the distance between the negative imaginary sources and the detector, where z0=1μs and zb is defined by

Eq. 4

zb=1+Reff1Reff2D1.
Following Ref. 48, Reff represents the fraction of photons that are internally diffusely reflected at the boundary and is assumed to be 0.493 for a refractive index of 1.4 in tissue.

An estimation of the absorption coefficient can be made using the Neumann boundary condition:36

Eq. 5

limtddtlogeR(ρ,t)=μac.
One can then use a linear regression on the tail of the logarithm of the reflectance to estimate μa .

2.2.

Two-Layered Medium

For a two-layer medium, the radiance is again given by Eq. 1, but this time Φ is the solution of the diffusion equation for a two-layer medium under the extrapolated boundary condition.22, 33, 34 The expression in the first layer is given in the Fourier domain by

Eq. 6

ϕ1(z,s,ω)=sinh[α1(zb+z0)]D1α1 ×D1α1cosh[α1(lz)]+D2α2sinh[α1(lz)]D1α1cosh[α1(l+zb)]+D2α2sinh[α1(l+zb)]sinh[α1(z0z)]D1α1for0zz0,
where α12=(D1s2+μa1+iωc)D1 , and s is the radial spatial frequency.

The spatial Fourier inversions, over s , must be done numerically, because no analytical solutions are available. To avoid numerical errors, hyperbolic functions in Eq. 6 are expanded. Because we assume cylindrical symmetry of the problem, the two-dimensional Fourier inversion of the preceding expression is given by

Eq. 7

Φω(ρ,s,ω)=A2π0dsϕ1(s,z,ω)sJ0(sρ),
where J0 is the Bessel function of zeroth order, and A is an amplitude factor as described previously. The Hankel transform in Eq. 7 is done numerically by using a Gauss-Laguerre quadrature of 5000 points. The nodes and the weights are calculated with MATLAB using the algorithm presented in Ref. 49. To obtain the TD reflectance, one must calculate the real and the imaginary part of

Eq. 8

R(ρ,ω)=0.118Φω(ρ,z=0,ω)+0.306Φωzz=0
for a set of frequencies, then perform a fast Fourier transform.

2.3.

Hemoglobin Concentration Recovery

We are interested in recovering hemoglobin concentrations. By assuming that oxy- and deoxyhemoglobin, as well as water, are the dominant absorbers between 690 and 850nm in tissue, and ignoring other background absorbers,32, 50, 51, 52 one can construct the following linear system:

Eq. 9

[μa(λ1)μa(λ2)μa(λ3)μa(λ4)]=[ϵHbO2(λ1)ϵHbR(λ1)ϵH2O(λ1)ϵHbO2(λ2)ϵHbR(λ2)ϵHO2(λ2)ϵHbO2(λ3)ϵHbR(λ3)ϵH2O(λ3)ϵHbO2(λ4)ϵHbR(λ4)ϵH2O(λ4)][CHbO2CHbRCH2O].
In this work, we assumed 70% of water for both the superficial tissues and the brain.32 The system was thus reduced to four equations and two unknowns and was overdetermined. The hemoglobin concentrations were recovered by inverting Eq. 9 with a least-square fit. The extinction coefficients ϵ(λ) can be found in the literature53, 54 while μa(λ) were recovered from the measurements.

3.

Methodology

A system consisting of four pulsed lasers (PicoQuant, Berlin, Germany) operating at 690, 750, 800, and 850nm , and four single-photon counting photomultiplier tubes (PMTs), was used to collect the light. A 4:1 combiner temporally interlaced the four wavelengths in a single fiber. Four detection fibers were placed at 10, 15, 25, and 30mm , respectively, from the source fiber on a rubber patch. The latter was then placed on the forehead of the subject. Measurements were taken for both the left and right hemisphere. Photons were collected for a period of 300seconds to generate sufficient statistics, i.e., a total count of more than 104 photons.

Five healthy adult subjects were recruited for this experiment. Every one previously passed a magnetic resonance imaging (MRI) exam (Siemens MAGNETOM Trio 3T, Malvern, Pennsylvania). We used the anatomical T1 image to measure the thickness of the scalp, skull, and cerebrospinal fluid (CSF) layers with the software SPM5 (UCL, London, UK). For the optical measures, the subjects were seated on a comfortable chair and were asked to stay calm during the acquisition. The experiment took place in a dark room to reduce the noise on the PMTs.

The fitting procedure used a nonlinear optimization routine (MATLAB function lsqcurvefit) with parameters μa , μs , and the amplitude factor A to get the best fit of the theoretical model to the experimental temporal point spread functions (TPSFs). The 16 TPSFs (four distances and four wavelengths) were first fit individually with a homogeneous model, Eqs. 1, 2, with three parameters: μa , μs , and the amplitude factor A . The TPSFs were then fit simultaneously four by four (four distances for each wavelength) using the two-layered model, Eqs. 6, 7, 8, using eight parameters ( μa and μs for each layer and an amplitude factor for each TPSF). For each subject, the initial values for these two-layered fits were taken to be the values obtained by the previous homogeneous fits at 10mm and the linear fit of the tail of the reflectance at 30mm for the first and second layer, respectively. The temporal range for the fit was from 20% of the maximum prior to the maximum peak, to 0.5% after the peak. Similar limits are found in the literature.55

The TPSFs measured with the apparatus were the convolutions of the TPSF due to the optical properties of the scattering medium with the instrument response function. Since each PMT had a characteristic response function, we measured each separately with a simple setup: the source and detector fibers faced each other and were separated by a thin piece of scattering paper to prevent PMT saturation and fill the whole numerical aperture of the detection fiber.56 In the fitting procedure, each theoretical TPSF function of the medium was convolved with the experimental instrument response function (IRF) before comparison with the experimental TPSF. When IRF measurements are done well, this procedure avoids including a jitter delay time in the fitting parameters.

Finally, hemoglobin concentrations were recovered by inverting Eq. 9 for each layer of the model.

4.

Results

4.1.

Simulation of the Recovery Procedure

The fitting procedure was first tested on four simulated data sets with different values for the first and the second layer. These sets were generated using a 100×100×100mm volume with 0.5×0.5×0.5mm voxels. The thickness of the first layer was 10mm , and the distance between the source and the detector was 30mm . The simulations were made using a three-dimensional Monte Carlo code.57 We simulated a peak amplitude of 104 photons, giving a photon noise of 1% at the peak, which is comparable to the experimental peak noise (<1%) . Results in Table 1 show less than an 11.6% error on the recovery of the absorption coefficient of the first layer, and less than a 3.1% error for the second layer, by using the multi-distance fitting procedure. In these recoveries, we assumed that the thickness of the first layer was known. Table 1 shows that the linear fit of the tail of the TPSF logarithm gave a better estimate of the absorption coefficient than using a homogeneous model.

Table 1

Absorption coefficients recovered with the two-layered model from four Monte Carlo data sets for both layers of the model. The thickness of the first layer was 10mm . Absorption coefficients were recovered from a single TPSF with source-detector distances of 10 and 30mm , and also from a multi-distance fit (10, 15, 25, and 30mm ). Results obtained using a homogeneous model are also shown for comparison.

Data set μa1 (mm−1) Error (%) μa2 (mm−1) Error (%) μa (mm−1)
RealRecoveredRealRecoveredHomogenousSlope
10mm
10.008090.00810.08940.01710.01549.530.01120.0204
20.008870.008079.010.01670.01510.50.01120.0203
30.008490.008530.4320.01720.015410.10.01170.0208
40.01010.01031.620.02040.016718.20.01370.0231
30mm
10.008090.008818.880.01710.01763.060.01320.0158
20.008870.008840.2840.01670.01733.070.01310.0156
30.008490.009248.870.01720.01762.610.01350.016
40.01010.01086.450.02040.02145.110.01560.0181
4 TPSFs
10.008090.0090311.60.01710.01653.09
20.008870.008890.2280.01670.01641.93
30.008490.009410.80.01720.01672.6
40.01010.01086.340.02040.02011.44

Separately, we simulated a complete Monte Carlo data set of 16 TPSFs, four source-detector distances (10, 15, 25, and 30mm ) for four different wavelengths (690, 750, 800, and 850nm ) with absorption coefficients computed from given hemoglobin concentrations in the superficial layer and in the brain. The concentrations chosen were 33 and 10μM in the superficial layer, and 65 and 22μM in the brain for HbO2 and HbR, respectively. The thickness of the first layer was also 10mm in these simulations. Recovered concentrations are shown in Fig. 1 . Using the two-layered model, the errors in the recovered concentrations were less than 10% and 2% in the superficial layer and in the brain, respectively, when the thickness of the first layer was known. However, as shown in Fig. 1, the homogeneous model underestimated the hemoglobin concentrations in the brain by a factor of 30%.

Fig. 1

Hemoglobin concentrations and SO2 recovered in the superficial layer and in the brain on simulated data.

054019_1_035805jbo1.jpg

4.2.

Thickness Sensitivity

To test the sensitivity of the procedure to the thickness of the first layer, hemoglobin concentrations were recovered from the same simulated data set used in the previous section but assuming different thicknesses, between 8 and 12mm , for the first layer. Errors in the HbO2 , HbR, and total hemoglobin (HbT) concentrations, and the oxygen saturation (SO2) recovered with this procedure, are shown in Fig. 2 . These results show that the accuracy of the recovery depends on the assumed thickness. For an error of ±1mm , the errors in the recovered hemoglobin concentrations in the brain were between 10% and 15%. On the other hand, the SO2 error stayed under 2%.

Fig. 2

Errors on the recovered concentrations related to the assumed thickness of the first layer. HbO2 , HbR, HbT, and SO2 errors are shown for data simulated with a first layer thickness of 10mm .

054019_1_035805jbo2.jpg

We also observed that the error in the hemoglobin concentrations in the brain due to an error in the estimation of the thickness of the first layer were always higher than the errors in the superficial layer. This was due to the brain’s low contribution to the TPSFs for source-detector separations between 1 and 3cm , so a small variation in the thickness resulted in a large relative variation in the brain’s contribution to the TPSFs. This also explains the fact that underestimating the thickness of the superficial layer resulted in smaller errors for the hemoglobin concentration in the brain, as shown in Fig. 2, compared to an overestimation. However, when the thickness of the first layer was well estimated, the errors were similar in the superficial layers and in the brain.

Figure 2 also shows the sign of the error in the recovered concentrations. For an overestimation of the first-layer thickness, the contribution of the brain to the TPSF decreased, which resulted in an overestimation of the brain’s hemoglobin to fit the theoretical model. As such, this overestimation produced an underestimation of hemoglobin in the superficial layer. On the other hand, if one underestimates the skin, skull, and CSF thicknesses, the contribution of the brain to the TPSF increases. This results in an underestimation of brain hemoglobin concentrations while superficial layers ones are less sensitive.

In practice, the thickness of the first layer enclosing skin, skull, and CSF can be determined very precisely using a high-resolution MRI scan. The error on the measurement is of the order of the voxel size. Here, the voxels were 1mm3 with a 3T magnet, meaning that the maximum errors on the hemoglobin concentrations were expected to be no more than 15%.

4.3.

In Vivo Measurements of Baseline Values

A typical example of the recovery procedure for in vivo data is presented in Fig. 3 . The first step consisted of recovering the thickness of the model’s first layer from an anatomical MRI image that comprised skin, skull, and CSF. After the fits were done with Eq. 1, the final step consisted of inverting Eq. 9 to recover hemoglobin concentrations. Detailed results for all subjects are presented in Table 2 for left-and right-side measurements on the forehead. For comparison, concentrations computed with absorption coefficients recovered with a homogeneous model are shown in Table 3 . A group average was also performed with concentrations recovered using the two-layered model, and results are presented in Table 4 .

Fig. 3

Typical procedure applied to recover the hemoglobin concentrations. (a) Recovery of the thickness of the first layer enclosing skin, scalp, and CSF on the MRI image. (b) Recovery of the reduced scattering and absorption coefficients for both layers using the fitting procedure. (c) Recovery of the chromophore concentrations with the inversion of Eq. 9.

054019_1_035805jbo3.jpg

Table 2

Hemoglobin concentrations measured on the left and right hemisphere for each subject computed with absorption coefficients recovered using a two-layered model.

Subject HbO2 (μM) HbR (μM) HbT (μM) SatO2 (%)
SuperficialBrainSuperficialBrainSuperficialBrainSuperficialBrain
1 left (2 layers)4.219.11.3315.95.533575.954.5
1 right (2 layers)2.0324.53.2516.45.2840.938.459.8
2 left (2 layers)5.8733.73.4239.2656.763.359.5
2 right (2 layers)4.3532.74.2620.58.653.250.561.5
3 left (2 layers)15.939.611.121.82761.458.963.7
3 right (2 layers)17.743.411.82029.663.46068.4
4 left (2 layers)27.135.613.926.4416266.157.4
4 right (2 layers)25.836.713.922.739.759.465.161.8
5 left (2 layers)10.421.25.481715.938.265.455.4
5 right (2 layers)12.723.74.4915.317.23973.960.8

Table 3

Hemoglobin concentrations measured on the left and right hemisphere for each subject computed with absorption coefficients recovered using a homogeneous model for the superficial layer and the linear fit of the TPSF slope for the brain. Superficial layer concentrations correspond to measurements at 10mm , while brain concentrations correspond to measurements at 30mm .

Subject HbO2 (μM) HbR (μM) HbT (μM) SO2 (%)
SuperficialBrainSuperficialBrainSuperficialBrainSuperficialBrain
1 left (1 layer)8.0822.53.1417.611.240.17256.1
1 right (1 layer)15.324.45.6817.22141.772.958.7
2 left (1 layer)22.931.55.9918.728.950.179.362.8
2 right (1 layer)19.931.67.71727.648.672.164.9
3 left (1 layer)19.216.98.2825.927.542.869.939.6
3 right (1 layer)2831.613.422.441.45467.758.5
4 left (1 layer)19.233.911.917.731.251.661.765.7
4 right (1 layer)23.839.721.718.345.45852.368.5
5 left (1 layer)1224.810.118.822.243.654.256.8
5 right (1 layer)12.225.97.8717.220.143.160.860

Table 4

Average hemoglobin concentrations for all the subjects. Concentrations were computed using the two-layered model.

Chromophore Left Right Average
SuperficialBrainSuperficialBrainSuperficialBrain
HbO2 12.729.812.532.212.631
(μM) ±9.25 ±9.12 ±9.77 ±8.31 ±8.97 ±8.32
HbR7.0520.87.54197.2919.9
(μM) ±5.3 ±4.34 ±4.92 ±3.05 ±4.83 ±3.67
HbT19.750.720.151.219.950.9
(μM) ±14.4 ±13 ±14.4 ±10.9 ±13.6 ±11.3
SO2 65.958.157.662.561.760.3
(%) ±6.24 ±3.66 ±13.6 ±3.43 ±10.9 ±4.06

5.

Discussion

Tables 2, 4 show a clear distinction between the superficial layer and brain for hemoglobin baseline concentrations (as expected). This distinction is in agreement with results recovered using FD system (Choi 8). The HbO2 and HbR concentrations were lower in the superficial layer than in the brain. The hemoglobin in the first layer was probably concentrated in the skin, which is highly irrigated with blood vessels; but this layer also included CSF, which contains low hemoglobin concentrations. However, our model did not validate this assumption. A three-layered model has been developed,42 but to our knowledge, no results were reported. Our hemoglobin concentrations, computed with the homogeneous model, were overestimated in the superficial layer—as expected from the Monte Carlo simulation. However, the difference was lower for brain hemoglobin estimations, because the linear fit of the TPSF slopes gave a better estimate of the absorption coeffcient than the fit with the homogeneous analytical expression, as shown in Table 1. Our results suggest that the layered structure of the head should be taken into account to extract hemoglobin concentrations for the brain and to confirm the need for using a two-layered model.

We also observed a large intersubject variability for hemoglobin concentrations, shown in Table 2. This fact was also reported both in Choi 8 and Comelli, 26 thus reinforcing the importance of baseline physiology measurements before interpreting activation maps of individual subjects. Simulation results from Table 2 and Fig. 2 show that the uncertainty in the estimated hemoglobin concentrations were less than 15% with our recovery procedure, which is due primarily to the estimation of the thickness of the superficial layer. This uncertainty alone cannot explain our observed variability between subjects. Differences in anatomical structure may explain this variability, but the subject’s baseline metabolism at the moment of the experiment may also play a role. Physical activity, for example, just before the measurement raises the blood circulation in the body, which may alter the hemoglobin concentration of the brain tissue. Moreover, the two-layered model does not separate the CSF layer from the skin and skull, which can cause changes in light distribution. Also, the model assumes that all the boundaries are parallel and flat, which is obviously not the case in reality. Finally, other background absorbers can contribute to absorption, but their contributions are considered very low according to the literature.32, 50, 51, 52

By looking at the SO2 variations in the brain, we see in Table 2 that these variations are only 4% between subjects. This result was also reported in Choi Figure 2 shows that the errors in the oxygen saturation are low (<2%) even if the assumption for the superficial layer thickness is bad. This makes sense, because the partial volume effect5 reduces the magnitude of the estimated concentration. However, with a good choice of wavelengths, we can reduce the crosstalk and preserve the relative magnitudes for HbO2 and HbR,58 which gives good SO2 values.

6.

Conclusion

We reported baseline hemoglobin concentrations measured in human intra- and extracerebral tissue using a TD system with a two-layered analytical model for the first time. We showed that hemoglobin concentrations recovered with TD systems differ when a two-layered model is used instead of a homogeneous one. Clear distinctions were obtained between the superficial layer (skin+skull+CSF) and the brain. A large intersubject variability was observed, as previously reported in the literature. The uncertainty for hemoglobin concentrations was estimated to be less than 15% for simulations dominated by uncertainty in the layer thickness, suggesting that the observed difference between subjects was real.

Acknowledgments

L. Gagnon is supported by a postgraduate scholarship from the Natural Sciences and Engineering Council of Canada (NSERC). This work was supported by the National Institutes of Health grants P41-RR14075 and R01-EB002482 and the NSERC Discovery grant.

References

1. 

A. Villringer, C. Hock, L. Schleinkofer, and U. Dirnagl, “Near infrared spectroscopy (NIRS): a new tool to study hemodynamic changes during activation of brain function in human adults,” Neurosci. Lett., 154 101 –104 (1993). https://doi.org/10.1016/0304-3940(93)90181-J 0304-3940 Google Scholar

2. 

Y. Hoshi and M. Tamura, “Multichannel near-infrared optical imaging of human brain activity,” J. Appl. Physiol., 75 1842 –1846 (1993). 8750-7587 Google Scholar

3. 

H. Obrig and A. Villringer, “Beyond the visible—imaging the human brain with light,” J. Cereb. Blood Flow Metab., 23 1 –18 (2002). https://doi.org/10.1097/00004647-200301000-00001 0271-678X Google Scholar

4. 

A. Gibson, J. Hebden, and S. Arridge, “Recent advances in diffuse optical imaging,” Phys. Med. Biol., 50 R1 –R43 (2005). https://doi.org/10.1088/0031-9155/50/4/R01 0031-9155 Google Scholar

5. 

D. Boas, T. Gaudette, G. Strangman, X. Cheng, J. Marota, and J. Mandeville, “The Accuracy of near infrared spectroscopy and imaging during focal changes in cerebral hemodynamics,” Neuroimage, 13 76 –90 (2001). 1053-8119 Google Scholar

6. 

M. Franceschini and D. Boas, “Noninvasive measurement of neuronal activity with near-infrared optical imaging,” Neuroimage, 21 372 –386 (2004). 1053-8119 Google Scholar

7. 

Y. S. Fawzi, A. B. M. Youssef, M. H. El-Batanony, and Y. M. Kadah, “Determination of the optical properties of a two-layer tissue model by detecting photons migrating at progressively increasing depths,” Appl. Opt., 42 6398 –6410 (2003). https://doi.org/10.1364/AO.42.006398 0003-6935 Google Scholar

8. 

J. Choi, M. Wolf, V. Toronov, U. Wolf, C. Polzonetti, D. Hueber, L. P. Safonova, R. Gupta, A. Michalos, W. Mantulin, “Noninvasive determination of the optical properties of adult brain: near-infrared spectroscopy approach,” J. Biomed. Opt., 9 221 –229 (2004). https://doi.org/10.1117/1.1628242 1083-3668 Google Scholar

9. 

H. Lim, J. F. de Boer, B. H. Park, E. C. Lee, R. Yelin, and S. H. Yun, “Optical frequency domain imaging with a rapidly swept laser in the 815870nm range,” Opt. Express, 14 5937 –5944 (2006). https://doi.org/10.1364/OE.14.005937 1094-4087 Google Scholar

10. 

A. Li, R. Kwong, A. Cerussi, S. Merritt, C. Hayakawa, and B. Tromberg, “Method for recovering quantitative broadband diffuse optical spectra from layered media,” Appl. Opt., 46 4828 –4833 (2007). https://doi.org/10.1364/AO.46.004828 0003-6935 Google Scholar

11. 

R. K. Wang and Y. A. Wickramasinghe, “Fast algorithm to determine optical properties of a turbid medium from time-resolved measurements,” Appl. Opt., 37 7342 –7351 (1998). 0003-6935 Google Scholar

12. 

H. Eda, I. Oda, Y. Ito, Y. Wada, Y. Oikawa, Y. Tsunazawa, M. Takada, Y. Tsuchiya, Y. Yamashita, M. Oda, “Multichannel time-resolved optical tomographic imaging system,” Rev. Sci. Instrum., 70 3595 –3602 (1999). https://doi.org/10.1063/1.1149965 0034-6748 Google Scholar

13. 

M. Schweiger and S. Arridge, “Application of temporal filters to time resolved data in optical tomography,” Phys. Med. Biol., 44 1699 –1717 (1999). https://doi.org/10.1088/0031-9155/44/7/310 0031-9155 Google Scholar

14. 

A. Pifferi, A. Torricelli, P. Taroni, and R. Cubeddu, “Reconstruction of absorber concentrations in a two-layer structure by use of multidistance time-resolved reflectance spectroscopy,” Opt. Lett., 26 1963 –1965 (2001). https://doi.org/10.1364/OL.26.001963 0146-9592 Google Scholar

15. 

F. Gao, Y. Tanikawa, H. Zhao, and Y. Yamada, “Semi-three-dimensional algorithm for time-resolved diffuse optical tomography by use of the generalized pulse spectrum technique,” Appl. Opt., 41 7346 –7358 (2002). https://doi.org/10.1364/AO.41.007346 0003-6935 Google Scholar

16. 

F. Martelli, S. D. Bianco, G. Zaccanti, A. Pifferi, A. Torricelli, A. Bassi, P. Taroni, and R. Cubeddu, “Phantom validation and in vivo application of an inversion procedure for retrieving the optical properties of diffusive layered media from time-resolved reflectance measurements,” Opt. Lett., 29 2037 –2039 (2004). https://doi.org/10.1364/OL.29.002037 0146-9592 Google Scholar

17. 

J. Selb, J. Stott, M. Franceschini, A. Sorenson, and D. Boas, “Improved sensitivity to cerebral hemodynamics during brain activation with a time-gated optical system: analytical model and experimental validation,” J. Biomed. Opt., 10 011013 (2005). https://doi.org/10.1117/1.1852553 1083-3668 Google Scholar

18. 

A. Torricelli, A. Pifferi, L. Spinelli, R. Cubeddu, F. Martelli, S. D. Bianco, and G. Zaccanti, “Time-resolved reflectance at null source-detector separation: improving contrast and resolution in diffuse optical imaging,” Phys. Rev. Lett., 95 078101 (2005). https://doi.org/10.1103/PhysRevLett.95.078101 0031-9007 Google Scholar

19. 

J. Selb, A. Dale, and D. Boas, “Linear 3D reconstruction of time-domain diffuse optical imaging differential data: improved depth localization and lateral resolution,” Opt. Express, 15 16400 –16412 (2007). https://doi.org/10.1364/OE.15.016400 1094-4087 Google Scholar

20. 

J. Swartling, J. Dam, and S. Andersson-Engels, “Comparison of spatially and temporal resolved diffuse-reflectance measurements systems for determination of biomedical optical properties,” Appl. Opt., 42 4612 –4620 (2003). https://doi.org/10.1364/AO.42.004612 0003-6935 Google Scholar

21. 

A. Liebert, H. Wabnitz, D. Grosenick, M. Mller, R. Macdonald, and H. Rinneberg, “Evaluation of optical properties of highly scattering media by moments of distribution of times of flight of photons,” Appl. Opt., 42 5785 –5790 (2003). https://doi.org/10.1364/AO.42.005785 0003-6935 Google Scholar

22. 

A. Kienle and T. Glanzmann, “In vivo determination of the optical properties of muscle with time-resolved reflectance using a layered model,” Phys. Med. Biol., 44 2689 –2702 (1999). https://doi.org/10.1088/0031-9155/44/11/301 0031-9155 Google Scholar

23. 

R. D. Hoge, M. A. Franceschini, R. J. M. Covolan, T. Huppert, J. B. Mandeville, and D. A. Boas, “Simultaneous recording of task-induced changes in blood oxygenation, volume, and flow using diffuse optical imaging and arterial spin-labeling MRI,” Neuroimage, 25 701 –707 (2005). 1053-8119 Google Scholar

24. 

R. D. Hoge, J. Atkinson, B. Gill, G. R. Crelier, S. Marrett, and G. B. Pike, “Investigation of BOLD signal dependence on cerebral blood flow and oxygen consumption:the deoxyhemoglobin dilution model,” Magn. Reson. Med., 42 849 –863 (1999). https://doi.org/10.1002/(SICI)1522-2594(199911)42:5<849::AID-MRM4>3.0.CO;2-Z 0740-3194 Google Scholar

25. 

A. N. Yaroslavsky, P. C. Schulze, I. V. Yaroslavsky, R. Schober, F. Ulrich, and H.-J. Schwarzmaier, “Optical properties of selected native and coagulated human brain tissues in vitro in the visible and near infrared spectral range,” Phys. Med. Biol., 47 2059 –2073 (2002). https://doi.org/10.1088/0031-9155/47/12/305 0031-9155 Google Scholar

26. 

D. Comelli, A. Bassi, A. Pifferi, P. Taroni, A. Torricelli, R. Cubeddu, F. Martelli, and G. Zaccanti, “In vivo time-resolved reflectance spectroscopy of the human forehead,” Appl. Opt., 47 1717 –1725 (2007). 0003-6935 Google Scholar

27. 

D. Contini, A. Torricelli, A. Pifferi, L. Spinelli, F. Paglia, and R. Cubeddu, “Multi-channel time-resolved system for functional near infrared spectroscopy,” Opt. Express, 14 5418 –5432 (2007). https://doi.org/10.1364/OE.14.005418 1094-4087 Google Scholar

28. 

R. B. Saager and A. J. Berger, “Direct characterization and removal of interfering absorption trends in two-layer turbid media,” J. Opt. Soc. Am. A, 22 1874 –1882 (2005). https://doi.org/10.1364/JOSAA.22.001874 0740-3232 Google Scholar

29. 

T. S. Leung, C. E. Elwell, and D. T. Delpy, “Estimation of cerebral oxy- and deoxy-haemoglobin concentration change in a layered adult head model using near-infrared spectroscopy and multivariate statistical analysis,” Phys. Med. Biol., 50 5783 –5798 (2005). https://doi.org/10.1088/0031-9155/50/24/002 0031-9155 Google Scholar

30. 

Q. Zhang, E. N. Brown, and G. E. Stangman, “Adaptative filtering for global interference cancellation and real-time recovery of evoked brain activity: a Monte Carlo simulation study,” J. Biomed. Opt., 12 044014 (2007). https://doi.org/10.1117/1.2754714 1083-3668 Google Scholar

31. 

Q. Zhang, E. N. Brown, and G. E. Stangman, “Adaptative filtering to reduce global interference in evoked brain activity detection: a human subject case study,” J. Biomed. Opt., 12 064009 (2007). https://doi.org/10.1117/1.2804706 1083-3668 Google Scholar

32. 

E. Ohmae, Y. Oouchi, M. Oda, T. Suzuki, S. Nobesawa, T. Kanno, E. Yoshikawa, M. Futatsubashi, Y. Ueda, H. Okada, “Cerebral hemodynamics evaluation by near-infrared time-resolved spectroscopy: correlation with simultaneous positron emission tomography measurements,” Neuroimage, 29 697 –705 (2006). 1053-8119 Google Scholar

33. 

A. Kienle, M. S. Patterson, N. Dognitz, R. Bays, G. Wagnieres, and H. van den Bergh, “Noninvasive determination of the optical properties of two-layered turbid media,” Appl. Opt., 37 779 –791 (1998). 0003-6935 Google Scholar

34. 

A. Kienle, T. Glanzmann, G. Wagnieres, and H. van den Bergh, “Investigation of two-layered turbid media with time-resolved reflectance,” Appl. Opt., 37 6852 –6862 (1998). 0003-6935 Google Scholar

35. 

F. Martelli, S. D. Bianco, and G. Zaccanti, “Procedure for retrieving the optical properties of a two-layered medium from time-resolved reflectance measurements,” Opt. Lett., 28 1236 –1238 (2003). 0146-9592 Google Scholar

36. 

M. S. Patterson, B. Chance, and C. Wilson, “Time resolved reflectance and transmittance for the noninvasive measurement of tissue optical properties,” Appl. Opt., 28 2331 –2336 (1989). 0003-6935 Google Scholar

37. 

J. M. Schmittt, G. X. Zhou, and E. C. Walker, “Multilayer model of photon diffusion in skin,” J. Opt. Soc. Am. A, 7 2141 –2153 (1990). 0740-3232 Google Scholar

38. 

I. Dayan, S. Havlin, and G. H. Weiss, “Photon migration in a two-layer turbid medium. A diffusion analysis,” J. Mod. Opt., 39 1567 –1582 (1992). 0950-0340 Google Scholar

39. 

D. Contini, F. Martelli, and G. Zaccanti, “Photon migration through a turbid slab described by a model based on diffusion approximation. I. Theory,” Appl. Opt., 36 4587 –4599 (1997). 0003-6935 Google Scholar

40. 

J. Ripoll, V. Ntziachristos, J. P. Culver, D. N. Pattanayak, A. G. Yodh, and M. Nieto-Vesperinas, “Recovery of optical parameters in multiple-layered diffusive media: theory and experiments,” J. Opt. Soc. Am. A, 18 821 –830 (2001). 0740-3232 Google Scholar

41. 

A. Sassaroli, F. Martelli, G. Zaccanti, and Y. Yamada, “Performance of fitting procedures in curved geometry for retrieval of the optical properties of tissue from time-resolved measurements,” Appl. Opt., 40 185 –197 (2001). 0003-6935 Google Scholar

42. 

F. Martelli, A. Sassaroli, Y. Yamada, and G. Zaccanti, “Analytical approximate solutions of the time-domain diffusion equation in layered slabs,” J. Opt. Soc. Am. A, 19 71 –80 (2002). 0740-3232 Google Scholar

43. 

F. Martelli, A. Sassaroli, S. Del-Bianco, Y. Yamada, and G. Zaccanti, “Solution of the time-dependent diffusion equation for layered diffusive media by the eigenfucntion method,” Phys. Rev. E, 67 056623 (2003). https://doi.org/10.1103/PhysRevE.67.056623 1063-651X Google Scholar

44. 

M. L. Shendeleva, “Time-domain Green functions for diffuse light in two adjoining turbid half-spaces,” Appl. Opt., 46 1641 –1649 (2007). https://doi.org/10.1364/AO.46.001641 0003-6935 Google Scholar

45. 

J. M. Tualle, H. L. Nghiem, D. Ettori, R. Sablong, E. Tinet, and S. Avrillier, “Asymptotic behavior and inverse problem in layered scattering media,” J. Opt. Soc. Am. A, 21 24 –34 (2004). https://doi.org/10.1364/JOSAA.21.000024 0740-3232 Google Scholar

46. 

A. Laidevant, A. da Silva, M. Berger, and J.-M. Dinten, “Effects of the surface boundary on the determination of the optical properties of a turbid medium with time-resolved reflectance,” Appl. Opt., 45 4756 –4764 (2006). https://doi.org/10.1364/AO.45.004756 0003-6935 Google Scholar

47. 

M. Shimada, Y. Hoshi, and Y. Yamada, “Simple algorithm for the measurement of absorption coefficient of a two-layered medium by spatially resolved and time-resolved reflectance,” Appl. Opt., 44 7554 –7562 (2005). https://doi.org/10.1364/AO.44.007554 0003-6935 Google Scholar

48. 

A. Kienle and M. S. Patterson, “Improved solutions of the steady-state and the time-resolved diffusion equations for reflectance from a semi-infinite turbid medium,” J. Opt. Soc. Am. A, 14 246 –254 (1997). https://doi.org/10.1364/JOSAA.14.000246 0740-3232 Google Scholar

49. 

G. H. Golub and J. H. Welsch, “Calculation of Gauss quadrature rules,” Math. Comput., 23 221 –230 (1969). https://doi.org/10.2307/2004418 0025-5718 Google Scholar

50. 

B. J. Tromberg, O. Coquoz, J. B. Fishkin, T. Pham, E. Anderson, J. Butler, M. Cahn, J. D. Gross, V. Venugopalan, and D. Pham, “Non-invasive measurements of breast tissue optical properties using frequency-domain photon migration,” Philos. Trans. R. Soc. London, Ser. B, 352 661 –668 (1997). https://doi.org/10.1098/rstb.1997.0047 0962-8436 Google Scholar

51. 

M. Cope, “The development of a near infrared spectroscopy system and its application for non invasive monitoring of cerebral blood and tissue oxygenation in the newborn infant,” University College London, (1991). Google Scholar

52. 

E. M. Sevick, B. Chance, J. Leigh, S. Nioka, and M. Maris, “Quantitation of time- and frequency-resolved optical spectra for the determination of tissue oxygenation,” Anal. Chem., 195 330 –351 (2007). 0003-2700 Google Scholar

53. 

S. Takatani and M. D. Graham, “Theoretical analysis of diffuse reflectance from a two-layer tissue model,” IEEE Trans. Biomed. Eng., 26 656 –664 (1979). 0018-9294 Google Scholar

54. 

G. M. Hale and M. R. Querry, “Optical constants of water in the 200nmto200m wavelength region,” Appl. Opt., 12 555 –563 (1973). https://doi.org/10.1007/BF00934777 0003-6935 Google Scholar

55. 

R. Cubeddu, A. Pifferi, P. Taroni, A. Torricelli, and G. Valentini, “Experimental test of theoretical models for time-resolved reflectance,” Med. Phys., 23 1625 –1633 (1996). https://doi.org/10.1118/1.597739 0094-2405 Google Scholar

56. 

A. Liebert, H. Wabnitz, D. Grosenick, and R. Macdonald, “Fiber dispersion in time domain measurements compromising the accuracy of determination of optical properties of strongly scattering media,” J. Biomed. Opt., 8 512 –516 (2003). https://doi.org/10.1117/1.1578088 1083-3668 Google Scholar

57. 

D. Boas, J. Culver, J. Stott, and A. Dunn, “Three-dimensional Monte Carlo code for photon migration through complex heterogenous media including the adult human head,” Opt. Express, 10 159 –170 (2002). 1094-4087 Google Scholar

58. 

G. Strangman, M. A. Franceschini, and D. A. Boas, “Factors affecting the accuracy of near-infrared spectroscopy concentration calculations for focal changes in oxygenation parameters,” Neuroimage, 18 865 –879 (2003). 1053-8119 Google Scholar
©(2008) Society of Photo-Optical Instrumentation Engineers (SPIE)
Louis Gagnon, Claudine Gauthier, Richard D. Hoge, Frederic Lesage, Juliette J. Selb, and David A. Boas "Double-layer estimation of intra- and extracerebral hemoglobin concentration with a time-resolved system," Journal of Biomedical Optics 13(5), 054019 (1 September 2008). https://doi.org/10.1117/1.2982524
Published: 1 September 2008
Lens.org Logo
CITATIONS
Cited by 55 scholarly publications and 7 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Brain

Absorption

Skin

Monte Carlo methods

In vivo imaging

Data modeling

Tissues

Back to Top