23 December 2020 Effects of lipid composition on photothermal optical coherence tomography signals
Author Affiliations +
Abstract

Significance: Photothermal optical coherence tomography (PT-OCT) has the promise to offer structural images coregistered with chemical composition information, which can offer a significant impact in early detection of diseases such as atherosclerosis.

Aim: We take the first step in understanding the relation between PT-OCT signals and the endogenous tissue composition by considering the interplay between the opto-thermo-physical properties of tissue as a function of its lipid composition and the ensuing effects on the PT-OCT signals.

Approach: Multiparameter theoretical estimates for PT-OCT signal as a function of composition in a two-component lipid–water model are derived and discussed. Experimental data from various concentrations of lipid in the form of droplets and injections under bovine cardiac muscle align with theoretical predictions.

Results: Theoretical and experimental results suggest that the variations of heat capacity and mass density with tissue composition significantly contribute to the amount of optical path length difference measured by OCT phase.

Conclusion: PT-OCT has the potential to offer key insights into the chemical composition of the subsurface lipid pools in tissue; however, the interpretation of results needs to be carried out by keeping the nonlinear interplay between the tissue of opto-thermo-physical properties and PT-OCT signals in mind.

Optical coherence tomography (OCT) is an interferometric optical imaging method which utilizes low-coherence light to form micrometer-resolution images of optical scattering media (e.g., biological tissues) in two and three dimensions.1 OCT systems are widely used in medical practice and research, specifically for interrogation of tissues that are incompatible with excisional biopsy (e.g., the eye, arteries, or nervous tissues).1,2 Conventional OCT measures the path length of elastically back-scattered light and provides images of tissue structure rather than chemical composition (scattering versus absorption of light). Consequently, OCT is quite sensitive to structural abnormalities induced at the early stages of diseases. However, in the cases where structural abnormalities are insufficient for clearly diagnosing the disease stage (e.g., atherosclerosis3,4 or early dental caries5), conventional OCT imaging often results in false positive readings, limiting the reliability and diagnostic value of OCT.

Photothermal optical coherence tomography (PT-OCT) is a functional extension of OCT with the promise to overcome the nonspecific nature of conventional OCT by forming three-dimensional (3D) images based on both scattering and absorption of light.6 In PT-OCT, an intensity-modulated photothermal (PT) laser with a wavelength in the absorption band of a molecule of interest is added to the conventional OCT system. In such a configuration, absorption of the PT laser by the molecule of interest induces a localized modulated temperature field (aka thermal wave field). This thermal wave field, in return, yields modulated thermo-elastic expansion resulting in modulated variation in the local refractive index. These phenomena ultimately lead to the modulation of the optical path length (OPL) at the modulation frequency of the PT laser with a modulation amplitude on the order of tens of nanometers. Since the OCT phase has enough sensitivity to measure such small variations in OPL, PT-OCT has the potential to offer coregistered structural and molecular information through the OCT signal’s amplitude and phase, respectively. To date, nanoparticles have been employed in PT-OCT as contrast agents to perform molecular-specific detection in human7 and rabbit8 tissues ex vivo as well as to detect cancer cells in vitro.9 More importantly, the feasibility of label-free imaging with PT-OCT has been demonstrated by measuring blood oxygen saturation in vessel phantoms,10 melanin distribution in zebra-fish eye,11 and lipids in human coronary arteries.12 Although these studies demonstrate the potential of PT-OCT for resolving and possibly quantifying molecules of interest in 3D, they assume a linear relation between the concentration of the molecule of interest and the PT-OCT signal. Although this assumption may be met for exogenous agents or molecules dissolved in aqueous media,10,11 in the more general case, variation in the composition of a tissue not only alters its optical properties but also induces changes in other physical properties (e.g., mass density and thermal properties). In this paper, we take the first step in understanding the relation between tissue composition and PT-OCT signals by considering the interplay between the opto-thermo-physical properties of tissue as a function of its composition. We develop a theoretical model estimating the PT-OCT response in a two-component tissue-like sample. We then present experimental PT-OCT results of mayonnaise (mayo)-ultrasound gel mixtures at various component ratios. Mayo was chosen to mimic the lipid-rich necrotic-core material present in atheromatous coronary atherosclerotic lesions.13,14 Mayo is primarily composed of lipids, which provide an absorption signature that can be targeted with PT-OCT, and its lipid composition is similar to that of atherosclerotic plaques.13,14 Since the composition and lipid content play a critical role in determining the propensity of a plaque to rupture,15 PT-OCT may offer the prospect to leverage the same contrast mechanism as that used by near-infrared spectroscopy16 and photoacoustic imaging1719 (i.e., absorption of lipid) for plaque composition imaging, albeit with much finer resolution.

Figure 1(a) shows schematic representation of processes that take place upon absorption of the PT laser in a PT-OCT system (e.g., absorption of 1210-nm laser by lipid–water mixture17). Absorption of intensity-modulated PT light results in production of heat Qin, which subsequently diffuses into the molecular matrix, forming a localized thermal wave field modeled by the bio-heat equation. When the PT laser spot size is small compared to the absorption depth (1/μa), the bio-heat equation is dominated by transverse/radial heat transfer.20 In such cases, the amplitude of temperature modulation, ΔT at a given depth and at the center of the PT beam is expressed by20.

Eq. (1)

ΔT=Pμa4απρcln(1+tLαω2/8),

Eq. (2)

power series;deg<2ΔT2PtLπω2×μaρc.
Here, P is the PT-laser power, μa is the medium’s absorption coefficient at the PT-laser wavelength, α is its thermal diffusivity, ρ is its mass density, c is its specific heat capacity, ω is the waist of the PT laser beam, and tL is the laser exposure time, taken as half of the PT laser modulation cycle, assuming a square modulation. By expanding the logarithm in Eq. (1) as a power series and neglecting terms of degree two and more, ΔT can be estimated as a simpler expression only containing the product of variables, Eq. (2). For a given PT-OCT setting, only parameters c, ρ, and μa in Eq. (2) are affected by the sample’s composition. Considering a linear variation of these parameters in a two-component mixture,21,22 the value of a given parameter can be modeled as

Eq. (3)

ζ=(ζ2ζ1)r+(ζ2+ζ1)2=ζranger+ζmean,0.5r0.5andζC=ζrangeζmean.
Here, ζ is the concentration-dependent material parameter (e.g., ρ), r is the relative concentration with respect to a 1:1 mixture, and ζ1, ζ2 are the values of the material parameters of pure individual components, respectively. ζC is the contrast of the material property in the two-component mixture and indicates the sensitivity of the parameter to variation of concentration. Using Taylor expansion, ΔT can then be expressed in terms of concentration-dependent material parameters as:

Eq. (4)

ΔT(r)=2PtLπω2×μmeanρmeancmean×(1+[μCρCcC]r+M(μC,ρC,cC)r2+).

Fig. 1

(a) Schematic presentation of sequence of physical processes taking place in PT-OCT upon absorption of PT light. (b) Numerical and experimental results for ΔOPL as a function of concentration of mayonnaise in the sample.

JBO_25_12_120501_f001.png

Equation (4) lies at the heart of this work and suggests that the temperature variation resulting from PT-laser illumination depends on nontrivial interactions of multiple physical parameters. Only when PT-OCT imaging is carried out on aqueous samples containing relatively low amounts of exogenous agents or dissolved molecules10,11 will ΔT depend linearly on their concentration. That is so because in these scenarios, changes in concentration do not lead to considerable change in density and specific heat capacity (i.e., ρCcC0), making the coefficients of nonlinear terms in Eq. (4) negligible. This linear prediction aligns well with the results reported to date on different concentrations of indocyanine green (ICG)-water mixtures23 or those on blood oxygen saturation.10 Variation of concentration in biological tissue compounds (e.g., lipid), on the other hand, results in considerable changes in density and specific heat capacity, leading to nonlinear variation of ΔT with concentration. The degree of this nonlinearity is determined by the contrasts of the component’s properties ζC.

Once ΔT is induced, the volume and the local refractive index of the molecular matrix change [Fig. 1(a)] and the optical path length difference (ΔOPL) as a function of relative concentration r can be found as24

Eq. (5)

ΔOPL(r)=L0L1n(T,r)dl=n(T0+ΔT)L1n(T0)L0=[n(r)+dndT(r)ΔT(r)][L0+L0β(r)ΔT(r)]n(r)L0ΔOPL(r)=L0{[n(r)β(r)+dndT(r)]ΔT(r)+[dndT(r)β(r)]ΔT(r)2},

Eq. (6)

ΔOPLΔT.

Here, T0 is the initial temperature of the sample, dn/dT is its thermo-optic coefficient, and β is its linear thermal expansion coefficient. Considering the range of values of parameters of Eq. (5) in biological systems (see Table 1), the quadratic term of ΔT is negligible compared to the linear term. Therefore, the ΔOPL measured through the phase signal of OCT is directly proportional to variations in temperature ΔT.

Table 1

Opto-thermo-physical properties of components used in simulations.

Property (unit)WaterMayo
Absorption coefficient μa (m1)1001716017
Density ρ (kg/m3)1000910
Specific heat c (J/kgK)4184245025,26
Thermal effusivity e (Ws1/2/m2K)158874021
Refractive index n1.34241.4927
Thermo-optic coefficient dndT91×10624531×10627
Linear thermal expansion coefficient β (K1)100×106241000×10628

To get a better understanding of the predictions of the developed model, we simulated the PT-OCT responses of various ratios of mayo–water mixtures. The opto-thermo-physical properties used for pure mayo and water (i.e., ζ1 and ζ2) are depicted in Table 1. The simulation results of the normalized ΔOPL as a function of mayo concentration are plotted in Fig. 1(b). In this plot, simulations are normalized with respect to the corresponding PT-OCT signals at 100% mayo concentration in order to make the results independent of the variation of PT laser parameters with depth (e.g., intensity or beam waist). When considering a mixture in which the properties of the solvent are dominant (ρCcC0), the change in the concentration predominantly induces a change in the absorption coefficient of the PT light. Under this condition, the nonlinear terms of Eq. (4) become negligible, resulting in a linear relation between ΔT [and consequently ΔOPL; Eq. (6)] and mayo concentration, as experimentally observed in solvent-dominant samples by other groups.10,23 However, when the sample is considered as a compound of mayo and water, the material properties are determined by the weight percentage of its components, Eq. (3). Since there is a pronounced difference between the opto-thermo-physical properties of mayo and water (Table 1), parameters ρc, cc, and μc in Eq. (4) are all considerable, leading to the nonlinear variation of ΔT with mayo concentration. Furthermore, the refractive index, its temperature dependence, and the expansion coefficient in Eq. (5) are also concentration-dependent (see Table 1) and contribute to the nonlinearity of ΔOPL seen in Fig. 1(b). Consequently, unlike the linear behaviors reported to date on samples with exogenous agents or dissolved molecules in aqueous media,10,11 the general variation of ΔOPL with the concentration of one tissue component is nonlinear. Another notable point in Fig. 1(b) is the small ΔOPL predicted for water (i.e., %mayo = 0) despite water’s moderate absorption at 1210 nm.17 The reason behind the small ΔOPL for water is the large heat capacity of water acting against the rise of the temperature as well as the cancellation of contributions of thermal expansion, n(r)β(r), and the thermo-optic coefficient, dndT(r), to OPL changes. The small ΔOPL of water is specifically important for PT-OCT of biological tissues that generally have high water content because it helps with the detection of PT-OCT signatures of waterless tissue components above the small PT-OCT signal baseline from water. Therefore, in PT-OCT, the large absorption coefficient of a component does not necessarily lead to a strong PT-OCT signal and, in fact, the optimization of PT-OCT experiments needs to take place by considering the interplay between the opto-thermo-physical properties of tissue.

To experimentally verify our theoretical predictions, we used a spectral-domain PT-OCT system (Fig. S1 in the Supplementary Material), employing broadband light of a superluminescent diode centered at 1310 nm (±75  nm at 10 dB; Exalos, Switzerland), a 2048-pixel line scan camera spectrometer with a maximum acquisition rate of 147  kHz (Wasatch Photonics; USA), and an intensity modulated single-mode PT laser (1210 nm; power on sample 18 mW). The axial and lateral resolutions of OCT in tissue were measured as 8.5 and 10  μm, respectively. Using a common-path reference, a displacement error of 3 nm at a signal-to-noise ratio of 35 dB was measured, which is close to that of a shot-noise-limited system.29 However, PT-OCT experiments were carried out without the phase reference using the absolute OCT phase (i.e., subject to inherent phase drifts of interferometer, laser scanning system, and other instabilities29). The OCT and PT beams are concentric and coscanned with an FWHM spot size of 8  μm. PT-OCT experiments were carried out at an A-line rate of 21.6 kHz. Each M-scan consisted of 5000 data points acquired in 230  ms. The PT signal was sinusoidal and modulated at 500 Hz, corresponding to the fundamental frequency of the square signal assumed for the theoretical derivation. The signal processing steps on a graphics processing unit (GPU) for forming PT-OCT images included: applying fast Fourier transformation (FFT) on the acquired spectra to get OCT amplitude and phase images, forming M-mode phase signals of each depth location, applying FFT on the M-mode phase signals for calculating PT-OCT image pixel amplitude using Eq. (7),30 removing outlier pixels, and subtracting base line

Eq. (7)

amp(Z)=|p(z,f0)|λ04π2f0Δt.
Here, |p| is the normalized FFT amplitude of the phase signal at the PT laser modulation frequency of f0, λ0 is the center wavelength of the OCT laser, and Δt is the acquisition time for one A-line.

To prepare various dilutions of the lipid–water compound, droplets of mayo and ultrasound gel with appropriate weight ratios were dispensed in a Petri dish and stirred well to make homogenous samples at five different weight concentrations of mayo (100%, 75%, 50%, 25%, and 0%). The scattering of the 0% mayo sample was adjusted to those of other mayo concentrations using titanium dioxide powder. For the dilution experiments, the average of the pixels in the PT-OCT B-mode images were normalized to that of the 100% mayo sample, as seen in Fig. 1(b). In addition, to examine the efficacy of the system in detecting lipid in tissue, lipid-rich atherosclerotic plaque was mimicked by injecting 95% mayo compound into fresh normal bovine cardiac muscle. A similar sample was also prepared using 40% mayo compound to examine the PT-OCT system’s sensitivity to lipid concentration in tissues. The thickness of the cap above the lipid pool for the 40% and 95% mayo simulated plaques varied between 110 and 360  μm and 140 to 280  μm, respectively. These simulated cap dimensions are comparable to those encountered in vulnerable coronary plaques.31

Figure 2 depicts results obtained from samples at various mayo concentrations. Conventional OCT images of (a)–(e) cannot discern between different concentrations of mayo/lipid. The root cause of this limitation is the fact that conventional OCT forms images based on elastic scattering of light, which originates from discontinuities in the materials’ refractive index rather than based on the presence of a characteristic molecular marker of the disease. Hence, tissues with very different compositions may have very similar OCT structural signatures. A case in point is the identification of thin-cap fibroatheromas, characterized by the presence of a necrotic core and extracellular lipid, and their differentiation from less vulnerable pathological intimal thickenings.3 Both plaque types bear similar structural features when imaged with intravascular OCT, complicating the identification of plaques at increased risk of causing future adverse events. The PT-OCT images of (f)–(j), on the other hand, clearly show the strong direct correlation of PT-OCT signal strength with concentration of mayo/lipid. The PT-OCT images of Fig. 2 qualitatively suggest a nonlinear variation of PT-OCT signal with concentration. The normalized average PT-OCT signals of each mayo concentration is plotted in Fig. 1(b) together with the simulated results. The nonlinear variation of PT-OCT signal with %mayo is evident in this plot, as predicted by the theoretical model. Another key observation is the significant underestimation of ΔOPL contrast between 0% and 100% mayo when only considering the absorption coefficient (μa) to be concentration-dependent in the model. In such a single-variable model, a 36-times higher absorption coefficient is needed for mayo to achieve the ΔOPL contrast seen in experimentation. The reason behind this discrepancy is that, in practice, both ρC and cC play an intensifying role based on Eq. (4) in producing a larger ΔT compared to the case where concentration dependence of ρC and cC are neglected. Also note that while the non-linear trend of variation of normalized ΔOPL with %mayo is reasonably close to the predicted theoretical values in the 25% to 100% range, the experimental PT-OCT signal deviates from the predicted theoretical trend at 0% (i.e., water) because the measured PT-OCT signal is dominated, in this case, by the noise floor in the OCT phase channel. The noise floor is likely limited by phase drift of the interferometer and could be reduced with a common-path reference. Using a fixed reflection signal in the sample arm just above the sample as a phase reference could also significantly reduce the phase noise, while being amenable to downstream translation into catheter-based systems (e.g., via phase reference taken from surface of the catheter ball lens).

Fig. 2

OCT images of samples with mayonnaise concentration: (a) 0%, (b) 25%, (c) 50%, (d) 75%, and (e)100%; PT-OCT images of samples with mayonnaise concentration (f) 0%, (g) 25%, (h) 50%, (i) 75%, and (j) 100%. Scale bars=50  μm.

JBO_25_12_120501_f002.png

Imaging results of the artificial lipid-rich plaques are shown in Fig. 3. In conventional OCT images, the boundary of the 40% mayo sample can clearly be identified in the OCT image of (a) while the boundary of the 95% mayo, (b), is diffuse, similar to those seen in OCT imaging of the lipid pool in human atherosclerotic plaques.32 In either case, information about the composition of the tissue cannot be obtained from the conventional OCT images. The PT-OCT images of (d) are from the sample with 95% mayo injection (i.e., similar lipid content as atherosclerotic plaques13,14). The lipid-rich pool is clearly identified in this image. The PT-OCT image of (c) is from the 40% mayo injection. Here, the subsurface diluted lipid pool cannot be reliably recognized as the PT-OCT signals are within the range of noise floor. While the lipid content here is 42% of that of the sample in (d), the resulting PT-OCT signal is only about 20% of that of the sample in (d). This nonlinear variation of PT-OCT signal with concentration is in complete alignment with the theoretical predictions of our model. Results of Fig. 3 indicate that PT-OCT has potential for offering key insights into the chemical composition of the subsurface lipid pools in tissue; however, the interpretation of results needs to be carried out with care and by keeping the nonlinear interplay between the opto-thermo-physical properties and PT-OCT signals in mind.

Fig. 3

OCT image of the plaque phantom with mayonnaise concentration: (a) 40% and (b) 95%; PT-OCT image of the plaque phantom with mayonnaise concentration (c) 40% and (d) 95% (scale bar=100  μm).

JBO_25_12_120501_f003.png

In summary, this work provides a theoretical model and a preliminary experimental study demonstrating the feasibility of obtaining insights into the chemical composition of biological tissues using PT-OCT. The developed model and experimental results suggest that the relation between PT-OCT signals and the concentration of an absorbing component in the tissue is nonlinear, and that the degree of this nonlinearity is determined by the contrasts of a component’s opto-thermo-physical properties ζC. Another key conclusion of this work is that in PT-OCT, a large absorption coefficient of a component does not necessarily lead to a strong PT-OCT signal. Instead, the strength of the PT-OCT signal is highly dependent on the interplay between the opto-thermo-physical properties of the tissue, all of which are concentration-dependent. Our experimental results suggest the possibility of obtaining insight into the lipid content of tissue-like samples and may offer a pathway toward refined assessment of atherosclerotic plaques in the future.

Disclosures

Authors have no relevant financial interests or conflicts of interest in this manuscript.

Acknowledgments

N. T. is grateful to the Natural Sciences and Engineering Research Council of Canada for the award of discovery grant (Grant No. RGPIN-2015-03666) as well as to the York University for the award of the Lassonde School of Engineering Innovation Fund. M. V. acknowledges support from the National Institutes of Health (Grant No. P41EB015903).

References

1. 

W. Drexler and J. G. Fujimoto, Optical Coherence Tomography: Technology and Applications, Springer International Publishing, Switzerland (2008). Google Scholar

2. 

B. E. Bouma et al., “Intravascular optical coherence tomography [Invited],” Biomed. Opt. Express, 8 (5), 2660 –2686 (2017). https://doi.org/10.1364/BOE.8.002660 BOEICL 2156-7085 Google Scholar

3. 

J. E. Phipps et al., “Diagnosis of thin-capped fibroatheromas in intravascular optical coherence tomography images: effects of light scattering,” Circ. Cardiovasc. Interventions, 9 (7), 10 (2016). https://doi.org/10.1161/CIRCINTERVENTIONS.115.003163 Google Scholar

4. 

G. J. Tearney et al., “Consensus standards for acquisition, measurement, and reporting of intravascular optical coherence tomography studies: a report from the international working group for intravascular optical coherence tomography standardization and validation,” J. Am. Coll. Cardiol., 59 (12), 1058 –1072 (2012). https://doi.org/10.1016/j.jacc.2011.09.079 JACCDI 0735-1097 Google Scholar

5. 

E. B. Shokouhi et al., “Comparative study on the detection of early dental caries using thermo-photonic lock-in imaging and optical coherence tomography,” Biomed. Opt. Express, 9 (9), 3983 –3997 (2018). https://doi.org/10.1364/BOE.9.003983 BOEICL 2156-7085 Google Scholar

6. 

D. C. Adler et al., “Photothermal detection of gold nanoparticles using phase-sensitive optical coherence tomography,” Opt. Express, 16 (7), 4376 –4393 (2008). https://doi.org/10.1364/OE.16.004376 OPEXFF 1094-4087 Google Scholar

7. 

C. Zhou et al., “Photothermal optical coherence tomography in ex vivo human breast tissues using gold nanoshells,” Opt. Lett., 35 (5), 700 –702 (2010). https://doi.org/10.1364/OL.35.000700 OPLEDP 0146-9592 Google Scholar

8. 

A. S. Paranjape et al., “Depth resolved photothermal OCT detection of macrophages in tissue using nanorose,” Biomed. Opt. Express, 1 (1), 2 –16 (2010). https://doi.org/10.1364/BOE.1.000002 BOEICL 2156-7085 Google Scholar

9. 

M. C. Skala et al., “Photothermal optical coherence tomography of epidermal growth factor receptor in live cells using immunotargeted gold nanospheres,” Nano Lett., 8 (10), 3461 –3467 (2008). https://doi.org/10.1021/nl802351p NALEFD 1530-6984 Google Scholar

10. 

R. V. Kuranov et al., “Depth-resolved blood oxygen saturation measurement by dual-wavelength photothermal (DWP) optical coherence tomography,” Biomed. Opt. Express, 2 (3), 491 –504 (2011). https://doi.org/10.1364/BOE.2.000491 BOEICL 2156-7085 Google Scholar

11. 

M. Lapierre-Landry et al., “Imaging melanin distribution in the zebrafish retina using photothermal optical coherence tomography,” Transl. Vision Sci. Technol., 7 (5), 4 (2018). https://doi.org/10.1167/tvst.7.5.4 Google Scholar

12. 

N. Katta et al., “Detection of lipid in coronary arteries using photothermal optical coherence tomography,” J. Am. Coll. Cardiol., 73 (9, Suppl. 1), 1421 (2019). https://doi.org/10.1016/S0735-1097(19)32027-3 JACCDI 0735-1097 Google Scholar

13. 

T. De Bruin et al., “Different postprandial metabolism of olive oil and soybean oil: a possible mechanism of the high-density lipoprotein conserving effect of olive oil,” Am. J. Clin. Nutr., 58 (4), 477 –483 (1993). https://doi.org/10.1093/ajcn/58.4.477 Google Scholar

14. 

C. P. Fleming et al., “Depth resolved detection of lipid using spectroscopic optical coherence tomography,” Biomed. Opt. Express, 4 (8), 1269 –1284 (2013). https://doi.org/10.1364/BOE.4.001269 BOEICL 2156-7085 Google Scholar

15. 

R. D. Madder et al., “Composition of target lesions by near-infrared spectroscopy in patients with acute coronary syndrome versus stable angina,” Circ. Cardiovasc. Interventions, 5 (1), 55 –61 (2012). https://doi.org/10.1161/CIRCINTERVENTIONS.111.963934 Google Scholar

16. 

R. D. Madder et al., “Detection by near-infrared spectroscopy of large lipid core plaques at culprit sites in patients with acute ST-segment elevation myocardial infarction,” JACC Cardiovasc. Interventions, 6 (8), 838 –846 (2013). https://doi.org/10.1016/j.jcin.2013.04.012 Google Scholar

17. 

T. J. Allen et al., “Spectroscopic photoacoustic imaging of lipid-rich plaques in the human aorta in the 740 to 1400 nm wavelength range,” J. Biomed. Opt., 17 (6), 061209 (2012). https://doi.org/10.1117/1.JBO.17.6.061209 JBOPFO 1083-3668 Google Scholar

18. 

K. Jansen et al., “Photoacoustic imaging of human coronary atherosclerosis in two spectral bands,” Photoacoustics, 2 (1), 12 –20 (2014). https://doi.org/10.1016/j.pacs.2013.11.003 Google Scholar

19. 

S. S. S. Choi et al., “Interference-free detection of lipid-laden atherosclerotic plaques by 3D co-registration of frequency-domain differential photoacoustic and ultrasound radar imaging,” Sci. Rep., 9 (1), 12400 (2019). https://doi.org/10.1038/s41598-019-48896-6 SRCEC3 2045-2322 Google Scholar

20. 

M. J. van Gemert, G. W. Lucassen and A. Welch, “Time constants in thermal laser medicine: II. Distributions of time constants and thermal relaxation of tissue,” Phys. Med. Biol., 41 (8), 1381 (1996). https://doi.org/10.1088/0031-9155/41/8/009 PHMBA7 0031-9155 Google Scholar

21. 

G. Szafner, D. Bicanic and O. Dóka, “Effect of fat content on the thermal effusivity of foods: an inverse photopyroelectric study,” Int. J. Food Prop., 14 (3), 666 –674 (2011). https://doi.org/10.1080/10942912.2010.498543 Google Scholar

22. 

A. Hagermann et al., “Physical properties as indicators of liquid compositions: derivation of the composition for Titan’s surface liquids from the Huygens SSP measurements,” Mon. Not. R. Astron. Soc., 359 (2), 637 –642 (2005). https://doi.org/10.1111/j.1365-2966.2005.08935.x MNRAA4 0035-8711 Google Scholar

23. 

M. Lapierre-Landry et al., “Photothermal optical coherence tomography of indocyanine green in ex vivo eyes,” Opt. Lett., 43 (11), 2470 –2473 (2018). https://doi.org/10.1364/OL.43.002470 OPLEDP 0146-9592 Google Scholar

24. 

A. Gupta, M. Villiger and N. Tabatabaei, “Optimal selection of laser modulation parameters in photothermal optical coherence tomography,” Proc. SPIE, 10067 1006718 (2017). https://doi.org/10.1117/12.2252779 PSISDG 0277-786X Google Scholar

25. 

X. Zhu, “Prediction of specific heat capacity of food lipids and foods,” (2015). Google Scholar

26. 

E. Contreras-Gallegos et al., “Specific heat of vegetable oils as a function of temperature obtained by adiabatic scanning calorimetry,” J. Therm. Anal. Calorim., 128 (1), 523 –531 (2017). https://doi.org/10.1007/s10973-016-5864-1 JTACF7 1418-2874 Google Scholar

27. 

H. Lim, “Temperature dependent refractive index of lipid tissue by optical coherence tomography imaging,” (2011). Google Scholar

28. 

A. Raudino et al., “Thermal expansion and compressibility coefficients of phospholipid vesicles: experimental determination and theoretical modeling,” J. Phys. Chem., 94 (10), 4217 –4223 (1990). https://doi.org/10.1021/j100373a063 JPCHAX 0022-3654 Google Scholar

29. 

E. W. Chang, J. B. Kobler and S. H. Yun, “Subnanometer optical coherence tomographic vibrography,” Opt. Lett., 37 (17), 3678 –3680 (2012). https://doi.org/10.1364/OL.37.003678 OPLEDP 0146-9592 Google Scholar

30. 

M. Lapierre-Landry, J. M. Tucker-Schwartz and M. C. Skala, “Depth-resolved analytical model and correction algorithm for photothermal optical coherence tomography,” Biomed. Opt. Express, 7 (7), 2607 –2622 (2016). https://doi.org/10.1364/BOE.7.002607 BOEICL 2156-7085 Google Scholar

31. 

J. Narula et al., “Histopathologic characteristics of atherosclerotic coronary disease and implications of the findings for the invasive and noninvasive detection of vulnerable plaques,” J. Am. Coll. Cardiol., 61 (10), 1041 –1051 (2013). https://doi.org/10.1016/j.jacc.2012.10.054 JACCDI 0735-1097 Google Scholar

32. 

H. Yabushita et al., “Characterization of human atherosclerosis by optical coherence tomography,” Circulation, 106 (13), 1640 –1645 (2002). https://doi.org/10.1161/01.CIR.0000029927.92825.F6 CIRCAZ 0009-7322 Google Scholar

Biography

Mohammadhossein Salimi received his BASc and MASc degrees in mechanical engineering from Tehran University and K.N.Toosi University of Technology, respectively. Currently, he is pursuing his PhD program in mechanical engineering with Hybrid Biomedical Optic Laboratory, York University, Toronto, Canada, focusing on optimizing Photo-Thermal Optical Coherence Tomography for detecting and characterizing lipid in biological tissues.

Martin Villiger received his MSc degree in microengineering and his PhD in photonics from the Swiss Federal Institute of Technology in Lausanne, Switzerland. Currently, he is an assistant professor (Harvard Medical School) at the Wellman Center for Photomedicine, located at the Massachusetts General Hospital in Boston, where he is leading research into advancing optical instrumentation for medical applications with an emphasis on polarization-sensitive and catheter-based optical coherence tomography.

Nima Tabatabaei is an associate professor of mechanical engineering at the York University (Canada). He received his PhD in mechanical engineering from the University of Toronto and carried out research at the Harvard Medical School/Wellman Center for Photomedicine as a postdoctoral fellow before joining the York University. His research interests are design and development of optical imaging technologies and point-of-need devices for early diagnosis and/or screening of diseases with an emphasis on the photothermal phenomenon.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Mohammad Hossein Salimi, Martin Villiger, and Nima Tabatabaei "Effects of lipid composition on photothermal optical coherence tomography signals," Journal of Biomedical Optics 25(12), 120501 (23 December 2020). https://doi.org/10.1117/1.JBO.25.12.120501
Received: 31 August 2020; Accepted: 1 December 2020; Published: 23 December 2020
Lens.org Logo
CITATIONS
Cited by 12 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Optical coherence tomography

Tissues

Absorption

Modulation

Water

Nonlinear optics

Tissue optics

Back to Top