23 March 2023 Timing resolution in double-sided silicon photon-counting computed tomography detectors
Author Affiliations +
Abstract

Purpose

Our purpose is to investigate the timing resolution in edge-on silicon strip detectors for photon-counting spectral computed tomography. Today, the timing for detection of individual x-rays is not measured, but in the future, timing information can be valuable to accurately reconstruct the interactions caused by each primary photon.

Approach

We assume a pixel size of 12 × 500 μm2 and a detector with double-sided readout with low-noise CMOS electronics for pulse processing for every pixel on each side. Due to the electrode width in relation to the wafer thickness, the induced current signals are largely dominated by charge movement close to the collecting electrodes. By employing double-sided readout electrodes, at least two signals are generated for each interaction. By comparing the timing of the induced current pulses, the time of the interaction can be determined and used to identify interactions that originate from the same incident photon. Using a Monte Carlo simulation of photon interactions in combination with a charge transport model, we evaluate the performance of estimating the time of the interaction for different interaction positions.

Results

Our simulations indicate that a time resolution of 1 ns can be achieved with a noise level of 0.5 keV. In a detector with no electronic noise, the corresponding time resolution is ∼0.1 ns.

Conclusions

Time resolution in edge-on silicon strip CT detectors can potentially be used to increase the signal-to-noise-ratio and energy resolution by helping in identifying Compton scattered photons in the detector.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Christel Sundberg, Mats U. Persson, Jacob Wikner, and Mats Danielsson "Timing resolution in double-sided silicon photon-counting computed tomography detectors," Journal of Medical Imaging 10(2), 023502 (23 March 2023). https://doi.org/10.1117/1.JMI.10.2.023502
Received: 7 February 2022; Accepted: 22 February 2023; Published: 23 March 2023
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
KEYWORDS
Electrodes

Silicon

Signal to noise ratio

Pulse signals

Photodetectors

Optical computing

Silicon photonics

RELATED CONTENT

Recent advances in silicon photonic integrated circuits
Proceedings of SPIE (February 13 2016)
A parametric design study of an electrochemical sensor
Proceedings of SPIE (August 21 2009)
Image Evaluation Of Image Intensifiers
Proceedings of SPIE (June 01 1974)

Back to Top