Open Access
18 August 2022 Biologically inspired intraoral camera for multifunctional dental imaging
Kisoo Kim, Kyung-Won Jang, Hyun-Kyung Kim, Sue Bean Cho, Ki-Hun Jeong
Author Affiliations +
Abstract

Biological vision offers intriguing inspiration for functional features in imaging systems with small form factors. We report biologically inspired intraoral camera (BIOC) for assorted dental imaging. This fully packaged BIOC features a convex-concave lens, inverted microlens arrays (iMLAs), LED module, and a single CMOS image sensor on a flexible printed circuit board in a handpiece holder. The iMLAs also collect light from wide angles by mounting the convex-concave lens to increase the viewing angle. The clinical trials have been successfully conducted for real-time and multifunctional intraoral monitoring of human teeth, including infinite depth of field, close-up, wide field-of-view, three-dimensional, and autofluorescence imaging. This biomedical camera provides insights for functional imaging not only in dental applications but also in surgical robots and endoscopy applications.

1.

Introduction

Biological vision in nature provides compelling opportunities for new functionality and small form factors in diverse imaging systems.17 In particular, about one million species of insects possess remarkable features such as wide viewing angle, large depth of field (DOF), low optical aberration, and polarization sensitivity.8,9 Compound eyes often exhibit a small number of retinula cells in a single channel, called ommatidium. In contrast, Xenos peckii has an unusual design of photoreceptors on an individual eyelet.10,11 Most compound eye cameras emulate their anatomical configuration of hemispherically arranged lenses and photoreceptors for large field-of-view (FOV) imaging.1214 However, the lens systems have technical limitations in low image resolution and complex fabrication processes. Artificial compound eye cameras inspired by Xenos peckii eye mainly used commercialized planar image sensors to comprise delicately defined multiple pixels in a single channel and achieve relatively high image resolution,7,1519 whereas the FOV of these type-cameras still remains narrow.

Inconvenient tools such as mirrors and cheek retractors are usually used for conventional dental examinations. The traditional dental tools have limitations that make it difficult for patients to directly check the condition of their teeth and require a wide opening of the mouth that causes discomfort to the patient. Compact intraoral dental cameras have been developed to overcome the discomfort and collect dental conditions compared with historical data.20,21 The intraoral cameras also offer several functions such as teeth surface reconstruction,2224 and dental health inspection.25 The dental camera accurately captures the structures and diseases of teeth through the compact camera integration and image reconstruction.26,27 For instance, the intraoral scanners often provide the anterior and posterior teeth arrangement as well as the position of the dental cavity.28,29 Dentists require teeth images from various angles such as right/left buccal and maxillary/mandibular occlusal for accurate diagnosis of teeth.30,31 However, previously developed intraoral cameras have technical bottlenecks such as large minimum object distance, thick total track length, narrow viewing angle, and the absence of functionality fusion.32,33

Here, we report an ultrathin intraoral camera for assorted functional dental imaging, inspired by the vision scheme of Xenos peckii’s eye, as shown in Fig. 1(a). The biologically inspired intraoral camera (BIOC) is fully packaged with a convex-concave lens, LED, and inverted microlens arrays (iMLAs) on a single complementary metal-oxide-semiconductor (CMOS) image sensor. The convex-concave lens collects light from wide-angle directions, coupled to the iMLAs, thus increases the viewing angle of BIOC about 1.7 times. A short focal length of iMLAs allows a wide DOF of BIOC for all-in-focus imaging, ranging from the surface of teeth to the palate. The fabrication steps of BIOC involve the microfabrication of iMLAs with optical block layers and the camera assembly, as shown in Fig. 1(b). Figure 1(c) shows the ultrathin array camera (UAC) packaged on a flexible printed circuit board (FPCB). The UAC, LED, and circuit components were mounted inside an anodized aluminum handpiece holder with 3-mm thickness to prevent deformation by sterilization. As shown in Figs. 1(d) and 1(e), the fully-assembled BIOC has a physical dimension of 7.8 mm in thickness and 25 mm in diameter and thus facilitates the intraoral operation for multifunctional dental imaging assembly. The thickness of BIOC is about 2.3 times as thin as that of a conventional wide-angle single-lens camera (WSLC, Raspberry Pi camera with 140-deg wide-angle lens, focal length: 2.72 mm).

Fig. 1

BIOC. (a) Schematic illustrations of BIOC, comprising a convex-concave lens, cover frames, LED, iMLAs, spacers, and CMOS image sensor on a FPCB. (b) Microfabrication steps and camera assembly of BIOC, using metal evaporation, lift-off, repetitive photolithography, and upside-down reflow. The cover frames were assembled through an epoxy adhesive for waterproofing. (c) The captured image of an UAC sensor packaged with the iMLAs and an image sensor mounted on the FPCB. (d) The comparison image of thickness between the BIOC and WSLC in the side-view. (e) Top view images of fully-assembled BIOC.

JOM_2_3_031202_f001.png

2.

Materials and Methods

The optical block layers consist of a 100-nm thick chromium (Cr) layer and double black SU-8 layers. LOR resist and AZ GXR-601 were patterned on a 4-inch borosilicate glass wafer, and the Cr film was deposited through evaporation on the patterns. The remaining resist patterns were removed through a lift-off process. A 5-μm thick black SU-8 (Gersteltec, GMC 1040) and a 25-μm thick SU-8 2025 were defined by repeating photolithography on the Cr patterns. AZ9260 resist was patterned into a cylindrical shape on the block layers, and an upside-down reflow process melted the cylinder patterns to form microlens shapes on a 180-deg convection oven. The fabricated wafer with the iMLAs was diced to 8  mm×8  mm, and the lens plate was packaged with an image sensor mounted on a flexible board (OmniVision OV5647, 5 M pixels, pixel size: 1.4  μm×1.4  μm) by using stacked alumina spacers with an epoxy adhesive. For the camera assembly, the convex-concave lens and LED were bonded on the cover frame by using UV curable adhesive (Norland, NOA63). The cover frames were combined using high shear and peel strength epoxy adhesive (3M, DP420) to prevent disinfectant penetration.

Multiple channel images captured by the BIOC were integrated through a high dynamic range imaging technique.34 The array images were cropped to each channel image, and the cropped images were aligned by matching the outer line of the images. The registered images were integrated by calculating a weighted average of each pixel. The weight factor was determined through the saturation and contrast of images.

3.

Results and Discussion

The DOF of BIOC was also characterized by a 3D dental phantom, as shown in Fig. 2(a). The experimental results show that the WSLC captures clear images at 3 cm from the central incisor of dental phantom but defocused images at close- or far distances over the DOF, for instance, 1 or 5 cm, respectively, as demonstrated in Fig. 2(b). The reason for image defocusing is because the image plane distance of WSLC rapidly transits according to the object distance, and the DOF of WSLC becomes shallow. Conventional intraoral cameras are often integrated with lens actuators to adjust a focusing plane, however, the actuator increases the total camera thickness and the actuator mounted camera still captures blurred images at a short distance less than about 1 cm.35 In contrast, the BIOC of a short focal length exceptionally exhibits a wide DOF, resulting in clear focused images at all object distances, as shown in Fig. 2(c). Figure 2(d) indicates that the BIOC clearly demonstrates a constant MTF50 along with the object distances, whereas the WSLC shows a significant decrease in MTF50 at object distances outside the DOF. As shown in Fig. 2(e), the array images of each channel acquire nearly similar scenes due to the wide viewing angle, thus multiple channel images were combined into a single image. The color difference of array images obtained from each channel is caused by the difference between the chief ray angles of the image sensor and that of iMLAs. Image reconstruction by merging multiple images slightly compensates for the color differences. In addition, a convex-concave lens on UAC substantially increases the FOV, as shown in Figs. 2(f) and 2(g). The BIOC captures the entire teeth of the 3D dental phantom due to a large FOV of 140-deg, whereas the UAC has the 80-deg in FOV, observing the partial teeth. The result indicates that the BIOC has a similar FOV to the WSLC with 140 deg. Note that the thickness of BIOC is 2.3 times smaller than that of WSLC.

Fig. 2

Dental phantom imaging through the WSLC and the BIOC. (a) Demonstration of phantom dental imaging through the fully-assembled BIOC. Optical images of the central incisor and the second molar were captured by (b) the WSLC and (c) the BIOC, depending on the object distances. The WSLC captures blurred images at close (1 cm) and far (5 cm) distances due to a limited depth-of-field. The BIOC provides clear images at all the different distances. (d) Comparison of MTF 50 between the captured images from the WSLC and the BIOC, which shows that dental images captured by the BIOC have a constant MTF value over the object distance. (e) Array image of dental phantom captured by the BIOC, and a reconstructed image through integral imaging. Optical images of the bottom dental arch were captured by (f) the UAC and (g) BIOC. The BIOC clearly observes the whole arrangement of the bottom dental arch compared to the UAC.

JOM_2_3_031202_f002.png

The BIOC delivers assorted functional images of human teeth after reconstructing captured array images. The human teeth imaging was performed after the KAIST institutional review board (IRB) approval. Three different cameras, including types A, B, and C, were compared for clinical trials, as shown in Fig. 3. The usability testing result shows that the A- and B-type intraoral cameras are convenient for observing the occlusal surface of teeth but inconvenient for investigating the inside of central incisors. The C-type camera is relatively suitable for observing all parts of teeth. For instance, it can capture the occlusal surface of second molars, anatomically located far away from incisors.

Fig. 3

Various types of intraoral camera prototypes for usability testing. Types A and B are slim intraoral cameras without the convex-concave lens and with the convex-concave lens, respectively. Type C is an intraoral camera with a dental mirror shape. The thickness of type A camera without the convex-concave lens is 5.5 mm, and that of types B and C is 7.8 mm.

JOM_2_3_031202_f003.png

The BIOC captures various human teeth images, such as the second molars, the second premolars, and the central incisors, as shown in Figs. 4(a)4(d). The second premolars are also captured from the buccal surface through a narrow gap between the cheek and the teeth. The integral imaging was conducted to improve color and hue contrast by combining array images into a single image. Figure 4(e) shows that the dental plaque image noticeably distinguishes the boundary of tooth and plaque after the image reconstruction of captured array images of BIOC. Acquired images were compared by expressing the color difference between a tooth, a gum, and plaque in a 1931 chromaticity diagram. The reconstructed image significantly enlarges the area of the color gamut by >2.23 times over a single-channel image, as demonstrated in Fig. 4(f). The experimental result also indicates that the color gamut area of the reconstructed image is comparable to that of WSLC, as shown in Fig. 4(g). The results indicate that the reconstructed image of BIOC has a color perception ability similar to that of a conventional single-lens camera image.

Fig. 4

Clinical trial for dental imaging through the BIOC. (a) Schematic illustration of dental arrangement for displaying the observation locations. Human dental images of (b) the second molar, (c) the second premolar, and (d) the central incisor were captured by the BIOC. (e) Human central incisor images are captured by (i) the single channel of BIOC (BIOC_S), (ii) reconstructed image from array images (BIOC_R), and (iii) the image of WSLC. (f) The CIE-1931 chromaticity diagram comparing color coordinates after the image reconstruction. The area of the color diagram is substantially increased by 2.23 times after the image reconstruction. (g) The color chromaticity diagram comparing images captured by the WSLC and the reconstructed image.

JOM_2_3_031202_f004.png

The visual disparity is observed between the anterior teeth’s top and bottom channel images in Fig. 5(a). The red-cyan anaglyph image is acquired by overlapping the top and bottom channel images as demonstrated in Fig. 5(b). The corresponding 3D color depth map exhibits the precise depth resolution of 84  μm, and the anterior teeth have a depth difference of 2 mm, as shown in Fig. 5(c). This depth disparity can be clinically used to determine malocclusion through the depth difference between the teeth of two arches.36 The autofluorescence imaging of BIOC further visualizes the plaque boundary under UV illumination. Oral bacteria or plaque can be observed with UV light due to the autofluorescence of porphyrin, i.e., organic compounds contained in dental biofilm, as shown in Fig. 5(d). The porphyrin emits red fluorescence when a UV light source is irradiated. The fluorescence image of dental plaque is captured through the BIOC, as demonstrated in Fig. 5(e). The fluorescence imaging of BIOC can be used to analyze the plaque area without errors clearly as well as widely observe the overall teeth. In addition, the dental inspection requires patient cooperation, such as widely opening the mouth. In particular, the maxillary and mandibular second molars, located distally from the incisors, have difficulty capturing clear images due to a lack of anatomical space.37,38 However, the BIOC overcomes the anatomical limitations through its thinness and allows capturing clear pictures with minimal discomfort.

Fig. 5

Multifunctional dental imaging. (a) Anterior dental images were extracted from the top and bottom channels of BIOC. (b) A red-cyan anaglyph image was acquired by merging the top and bottom channel images. (c) Reconstructed 3D depth color map of anterior teeth, showing the depth resolution of 84  μm as well as the depth difference of 2 mm between the top and bottom dental arches. (d) The principle of fluorescence dental imaging. The red fluorescence is emitted by autofluorescence of porphyrin when the teeth with porphyrin compound are irradiated with a UV light source. (e) A captured fluorescence dental image was recorded by the BIOC with a UV light source.

JOM_2_3_031202_f005.png

4.

Conclusions

We have successfully demonstrated a multifunctional intraoral camera inspired by the vision scheme of the Xenos peckii eye. The BIOC features the convex-concave lens, the cover frames, the LED, and the ultrathin camera. The BIOC allows all-in-focus dental imaging and multifunctional imaging by reconstructing captured array images. The experimental results exhibit that reconstructed images clearly increase the area of color space by 2.23 times and offer a 3D depth map with 84-μm depth accuracy. Also, the UV fluorescence dental imaging of BIOC clearly visualizes dental plaques. This biomedical camera module provides opportunities for exploring multifunctional imaging of diverse anatomical structures or epidermal surfaces through narrow natural orifices of humans.

Acknowledgments

This work was supported by the National Research Foundation of Korea (NRF) funded by the Ministry of Science ICT & Future Planning (2021R1A2B5B03002428, NRF-2021R1C1C1011567) and Ministry of Trade, Industry & Energy, Republic of Korea (P0013915). The authors declare no competing financial interests.

References

1. 

K. H. Jeong, J. Kim and L. P. Lee, “Biologically inspired artificial compound eyes,” Science, 312 (5773), 557 –561 (2006). https://doi.org/10.1126/science.1123053 SCIEAS 0036-8075 Google Scholar

2. 

Y. H. Jung et al., “Bioinspired electronics for artificial sensory systems,” Adv. Mater., 31 (34), 1803637 (2019). https://doi.org/10.1002/adma.201803637 ADVMEW 0935-9648 Google Scholar

3. 

T. Chung et al., “Mining the smartness of insect ultrastructures for advanced imaging and illumination,” Adv. Funct. Mater., 28 (24), 1705912 (2018). https://doi.org/10.1002/adfm.201705912 AFMDC6 1616-301X Google Scholar

4. 

M. S. Kim et al., “An aquatic-vision-inspired camera based on a monocentric lens and a silicon nanorod photodiode array,” Nat. Electron., 3 (9), 546 –553 (2020). https://doi.org/10.1038/s41928-020-0429-5 NEREBX 0305-2257 Google Scholar

5. 

J. J. Kim et al., “Biologically inspired artificial eyes and photonics,” Rep. Progr. Phys., 83 (4), 047101 (2020). https://doi.org/10.1088/1361-6633/ab6a42 Google Scholar

6. 

G. J. Lee et al., “Bioinspired artificial eyes: optic components, digital cameras, and visual prostheses,” Adv. Funct. Mater., 28 (24), 1705202 (2018). https://doi.org/10.1002/adfm.201705202 AFMDC6 1616-301X Google Scholar

7. 

S. I. Bae et al., “High contrast ultrathin light-field camera using inverted microlens arrays with metal–insulator–metal optical absorber,” Adv. Opt. Mater., 9 (6), 2001657 (2021). https://doi.org/10.1002/adom.202001657 2195-1071 Google Scholar

8. 

D. G. Stavenga and R. C. Hardie, Facets of Vision, Springer-Verlag Berlin Heidelberg(1989). Google Scholar

9. 

E. Warrant and D.-E. Nilsson, Invertebrate Vision, Cambridge University Press(2006). Google Scholar

10. 

E. Buschbeck, B. Ehmer and R. Hoy, “Chunk versus point sampling: visual imaging in a small insect,” Science, 286 (5442), 1178 –1180 (1999). https://doi.org/10.1126/science.286.5442.1178 SCIEAS 0036-8075 Google Scholar

11. 

D. Keum, H. Jung and K. H. Jeong, “Planar emulation of natural compound eyes,” Small, 8 (14), 2169 –2173 (2012). https://doi.org/10.1002/smll.201200107 SMALBC 1613-6810 Google Scholar

12. 

Y. M. Song et al., “Digital cameras with designs inspired by the arthropod eye,” Nature, 497 (7447), 95 –99 (2013). https://doi.org/10.1038/nature12083 Google Scholar

13. 

D. Floreano et al., “Miniature curved artificial compound eyes,” Proc. Natl. Acad. Sci. U. S. A., 110 (23), 9267 –9272 (2013). https://doi.org/10.1073/pnas.1219068110 Google Scholar

14. 

C. C. Huang et al., “Large-field-of-view wide-spectrum artificial reflecting superposition compound eyes,” Small, 10 (15), 3050 –3057 (2014). https://doi.org/10.1002/smll.201400037 SMALBC 1613-6810 Google Scholar

15. 

D. Keum et al., “Xenos peckii vision inspires an ultrathin digital camera,” Light: Sci. Appl., 7 (1), 1 –7 (2018). https://doi.org/10.1038/s41377-018-0081-2 Google Scholar

16. 

K. Kim et al., “Biologically inspired ultrathin arrayed camera for high-contrast and high-resolution imaging,” Light: Sci. Appl., 9 (1), 28 (2020). https://doi.org/10.1038/s41377-020-0261-8 Google Scholar

17. 

A. Bruckner et al., “Thin wafer-level camera lenses inspired by insect compound eyes,” Opt. Express, 18 (24), 24379 –24394 (2010). https://doi.org/10.1364/OE.18.024379 OPEXFF 1094-4087 Google Scholar

18. 

K. Kim et al., “Ultrathin arrayed camera for high-contrast near-infrared imaging,” Opt. Express, 29 (3), 1333 –1339 (2021). https://doi.org/10.1364/OE.409472 OPEXFF 1094-4087 Google Scholar

19. 

K. W. Jang et al., “Biologically inspired ultrathin contact imager for high-resolution imaging of epidermal ridges on human finger,” Adv. Mater. Technol., 6 (8), 2100090 (2021). https://doi.org/10.1002/admt.202100090 Google Scholar

20. 

S. Logozzo et al., “A comparative analysis of intraoral 3D digital scanners for restorative dentistry,” Internet J. Med. Technol., 5 (1), 1 –2 (2011). https://doi.org/10.5580/1b90 Google Scholar

21. 

M. Fujimoto et al., “Endoscopic system based on intraoral camera and image processing,” IEEE Trans. Biomed. Eng., 66 (4), 1026 –1033 (2018). https://doi.org/10.1109/TBME.2018.2866273 IEBEAX 0018-9294 Google Scholar

22. 

R. Richert et al., “Intraoral scanner technologies: a review to make a successful impression,” J. Healthcare Eng., 2017 8427595 (2017). https://doi.org/10.1155/2017/8427595 Google Scholar

23. 

M. Stanley et al., “Fully digital workflow, integrating dental scan, smile design and CAD-CAM: case report,” BMC Oral Health, 18 (1), 1 –8 (2018). https://doi.org/10.1186/s12903-018-0597-0 Google Scholar

24. 

P. Hong-Seok and S. Chintal, “Development of high speed and high accuracy 3D dental intra oral scanner,” Proc. Eng., 100 1174 –1181 (2015). https://doi.org/10.1016/j.proeng.2015.01.481 Google Scholar

25. 

F. Shakibaie and L. J. Walsh, “Dental calculus detection using the VistaCam,” Clin. Exp. Dent. Res., 2 (3), 226 –229 (2016). https://doi.org/10.1002/cre2.42 Google Scholar

26. 

N. Gan, Y. Y. Xiong and T. Jiao, “Accuracy of intraoral digital impressions for whole upper jaws, including full dentitions and palatal soft tissues,” PLoS One, 11 (7), e0158800 (2016). https://doi.org/10.1371/journal.pone.0158800 POLNCL 1932-6203 Google Scholar

27. 

J. Kuhnisch et al., “In vivo validation of near-infrared light transillumination for interproximal dentin caries detection,” Clin. Oral Invest., 20 (4), 821 –829 (2016). https://doi.org/10.1007/s00784-015-1559-4 Google Scholar

28. 

R. S. Jones et al., “Near-infrared transillumination at 1310-nm for the imaging of early dental decay,” Opt. Express, 11 (18), 2259 –2265 (2003). https://doi.org/10.1364/OE.11.002259 OPEXFF 1094-4087 Google Scholar

29. 

Y. Zhou et al., “Towards AR-assisted visualisation and guidance for imaging of dental decay,” Healthc. Technol. Lett., 6 (6), 243 –248 (2019). https://doi.org/10.1049/htl.2019.0082 Google Scholar

30. 

I. Ahmad, “Digital dental photography. Part 8: intra-oral set-ups,” Br. Dental J., 207 (4), 151 –157 (2009). https://doi.org/10.1038/sj.bdj.2009.715 Google Scholar

31. 

J. R. Dunn, “Dental photography: a new perspective,” Oral Health Pract. Manage., 6 (3), 34 –40 (2009). Google Scholar

32. 

J. S. Ahn et al., “Development of three-dimensional dental scanning apparatus using structured illumination,” Sens.-Basel, 17 (7), 1634 (2017). https://doi.org/10.3390/s17071634 Google Scholar

33. 

J. Kuhnisch et al., “Evaluation of detecting proximal caries in posterior teeth via visual inspection, digital bitewing radiography and near-infrared light transillumination,” Am. J. Dent., 32 (3), 74 –80 (2019). Google Scholar

34. 

T. Mertens, J. Kautz and F. Van Reeth, “Exposure fusion,” in 15th Pacific Conf. Comput. Graph. and Appl. (PG’07), 382 –390 (2007). https://doi.org/10.1109/PG.2007.17 Google Scholar

35. 

W.-S. Sun et al., “Simulation of autofocus lens design for a cell phone camera with object distance from infinity to 9.754 mm,” Appl. Opt., 54 (28), E203 –E209 (2015). https://doi.org/10.1364/AO.54.00E203 APOPAI 0003-6935 Google Scholar

36. 

B. Bell, “Paul G. Spencer,” Am. J. Orthodont., 51 (9), 693 –694 (1965). https://doi.org/10.1016/0002-9416(65)90262-9 Google Scholar

37. 

F. Mangano et al., “Intraoral scanners in dentistry: a review of the current literature,” BMC Oral Health, 17 (1), 1 –11 (2017). https://doi.org/10.1186/s12903-017-0442-x Google Scholar

38. 

J. Winkler and N. Gkantidis, “Trueness and precision of intraoral scanners in the maxillary dental arch: an in vivo analysis,” Sci. Rep.-UK, 10 (1), 1 –11 (2020). https://doi.org/10.1038/s41598-020-58075-7 Google Scholar

Biography

Kisoo Kim received his PhD at the Department of Bio and Brain Engineering from KAIST, in 2021. Now, he is a senior researcher in the Intelligent Optical Module Research Center at KOPTI. His research interests include optical module design and evaluation for compact arrayed camera system.

Kyung-Won Jang received his MS degree in bio and brain engineering from KAIST in 2015. Currently, he is a PhD candidate in bio and brain engineering at KAIST. His research interests include ultrathin MEMS cameras for healthcare applications.

Hyun-Kyung Kim received her MS degree in the Department of Bio and Brain Engineering from KAIST in 2021. Now, she is a PhD candidate in bio and brain engineering at KAIST. Her research interests include MLA camera for healthcare applications.

Sue Bean Cho received her MS degree in the College of Dentistry from YONSEI University in 2013. Now, she is a dentist at KAIST Clinic Pappalardo Center.

Ki-Hun Jeong received his PhD from the University of California at Berkeley in 2005. Currently, he is a full professor and the department head in the Department of Bio and Brain Engineering at KAIST, as well as a codirector at KAIST Institute for Health Science and Technology. His research focuses biophotonic MEMS/NEMS including bioinspired photonic devices, and materials for advanced clinical endoscopic imaging.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Kisoo Kim, Kyung-Won Jang, Hyun-Kyung Kim, Sue Bean Cho, and Ki-Hun Jeong "Biologically inspired intraoral camera for multifunctional dental imaging," Journal of Optical Microsystems 2(3), 031202 (18 August 2022). https://doi.org/10.1117/1.JOM.2.3.031202
Received: 7 April 2022; Accepted: 3 June 2022; Published: 18 August 2022
Advertisement
Advertisement
KEYWORDS
Cameras

Teeth

Imaging arrays

3D image reconstruction

Image sensors

Light emitting diodes

Luminescence

Back to Top