24 July 1998 Selenium direct-converter structure for static and dynamic x-ray detection in medical imaging applications
Brad T. Polischuk, Ziad Shukri, Anne Legros, Henri Rougeot
Author Affiliations +
Abstract
Digital x-ray imaging techniques of today require electronic detectors that can be applied to all modalities of medical imaging. This paper presents work showing that selenium, when used as a direct converter, can be competitive with present day scintillator technologies targeting mammographic, radiographic and fluoroscopic applications. In this work, supporting results are presented on the dark currents, lag and ghosting effects and also on the theoretical and experimental x-ray absorption and sensitivity of selenium layers. Measurements were carried out on suitably alloyed selenium layers where the electronic transport properties have been optimized. Measured values for dark currents were below 100 pA/cm2 at operating fields up to 20 V/micrometer. Experimental measurement of the intrinsic lag in selenium has shown it to be less than 0.5% after 30 milliseconds under a dose of 50 mR at 55 keV mean beam energy, which is very low compared with present day image intensifiers. Similar measurements on ghosting, using multiple radiographic pulses, indicate that the magnitude of the ghost image after a few seconds is around 2000 electrons, which is comparable to the electronic noise of most read out systems. Measured sensitivity of a 200 micrometer selenium layer under a mammographic spectrum was around 230 pC/mR/cm2 at an operating field of 20 V/micrometer, which is significantly higher than that reported for competing technologies. Sensitivity for 1000 micrometer selenium was also measured with an 80 kVp spectrum and 20 mm Al filtration and was found to be around 3400 pC/mR/cm2 which is in close agreement with theoretically calculated values. Theoretical estimations for MTF and DQE are also given to assess the potential imaging performance of a selenium-based detector for various applications.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Brad T. Polischuk, Ziad Shukri, Anne Legros, and Henri Rougeot "Selenium direct-converter structure for static and dynamic x-ray detection in medical imaging applications", Proc. SPIE 3336, Medical Imaging 1998: Physics of Medical Imaging, (24 July 1998); https://doi.org/10.1117/12.317051
Lens.org Logo
CITATIONS
Cited by 48 scholarly publications and 3 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Selenium

X-rays

Sensors

Absorption

X-ray detectors

Electrons

Metals

Back to Top