2 May 2008 Influences on accuracy of SEM based CD mask metrology with a view to the 32 nm node
W. Häßler-Grohne, C. G. Frase, D. Gnieser, H. Bosse, J. Richter, A. Wiswesser
Author Affiliations +
Proceedings Volume 6792, 24th European Mask and Lithography Conference; 67920O (2008) https://doi.org/10.1117/12.798783
Event: 24th European Mask and Lithography Conference, 2008, Dresden, Germany
Abstract
Scanning electron microscopy (SEM) is used today and will be used also in the near future as a fast and high resolution measurement method capable to perform characterizations of the smallest isolated and dense features which are to be specified and produced on photomasks down to the 32 nm node and perhaps below. It has been demonstrated however, that SEM based CD metrology results on different mask absorber stacks may show systematic dependencies on the absorber materials and other stack parameters between different CD-SEM equipment. These mask dependent effects have to be taken into account for proper cross-correlation of different CD metrology tools, which is required in mask manufacturing environment. In this contribution we will report on systematic investigations of simulated and measured SEM profiles of line and space features present on current and future mask generations as well as on suitable test structures. The investigations comprise variations of SEM primary energy settings and the response of different edge operators on these variations as well as variations of edge angle of the mask features. The simulations are based on Monte Carlo methods which are supplemented by additional phenomenological approaches where appropriate. Simulations of SEM edge contrasts and the response of different CD-SEM edge algorithms will be described for feature sizes down to 32 nm, because the smallest assist-features on the mask which have to be controlled with respect to their dimensions are about the same size as the smallest feature sizes on the wafer.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
W. Häßler-Grohne, C. G. Frase, D. Gnieser, H. Bosse, J. Richter, and A. Wiswesser "Influences on accuracy of SEM based CD mask metrology with a view to the 32 nm node", Proc. SPIE 6792, 24th European Mask and Lithography Conference, 67920O (2 May 2008); https://doi.org/10.1117/12.798783
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Photomasks

Critical dimension metrology

Scanning electron microscopy

Electrons

Monte Carlo methods

Calibration

Metrology

Back to Top