19 March 2008 Analysis of in-plane signal-to-noise ratio in computed tomography
Takanori Hara, Katsuhiro Ichikawa, Shigeru Sanada, Yoshihiro Ida
Author Affiliations +
Abstract
The purposes of this study are to analyze signal-to-noise ratio (SNR) changes for in-plane (axial plane) position and in-plane direction in X-ray computed tomography (CT) system and to verify those visual effects by using simulated small low-contrast disc objects. Three-models of multi detector-row CT were employed. Modulation transfer function (MTF) was obtained using a thin metal wire. Noise power spectrum (NPSs) was obtained using a cylindrical water phantom. The measurement positions were set to center and off-centered positions of 64mm, 128mm and 192mm. One-dimensional MTFs and NPSs for the x- and y-direction were calculated by means of a numerical slit scanning method. SNRs were then calculated from MTFs and NPSs. The simulated low-contrast disc objects with diameter of 2 to 10mm and contrast to background of 3.0%, 4.5% and 6.0% were superimposed on the water phantom images. Respective simulated objects in the images are then visually evaluated in degree of their recognition, and then the validity of the resultant SNRs are examined. Resultant in-plane SNRs differed between the center and peripheries and indicated a trend that the SNR values increase in accordance with distance from the center. The increasing degree differed between x- and y-direction, and also changed by the CT systems. These results suggested that the peripheries region has higher low-contrast detectability than the center. The properties derived in this study indicated that the depiction abilities at various in-plane positions are not uniform in clinical CT images, and detectability of the low contrast lesion may be influenced.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Takanori Hara, Katsuhiro Ichikawa, Shigeru Sanada, and Yoshihiro Ida "Analysis of in-plane signal-to-noise ratio in computed tomography", Proc. SPIE 6913, Medical Imaging 2008: Physics of Medical Imaging, 691353 (19 March 2008); https://doi.org/10.1117/12.768968
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Signal to noise ratio

Modulation transfer functions

Computed tomography

X-ray computed tomography

Spatial frequencies

Sensors

Visibility

Back to Top