22 March 2016 A wide-acceptance Compton spectrometer for spectral characterization of a medical x-ray source
Michelle A. Espy, A. Gehring, A. Belian, T. Haines, J. Hunter, M. James, M. Klasky, J. Mendez, D. Moir, R. Sedillo, R. Shurter, J. Stearns, K. Van Syoc, P. Volegov
Author Affiliations +
Abstract
Accurate knowledge of the x-ray spectra used in medical treatment and radiography is important for dose calculations and material decomposition analysis. Indirect measurements via transmission through materials are possible. However, such spectra are challenging to measure directly due to the high photon fluxes. One method of direct measurement is via a Compton spectrometer (CS) method. In this approach, the x-rays are converted to a much lower flux of electrons via Compton scattering on a converter foil (typically beryllium or aluminum). The electrons are then momentum selected by bending in a magnetic field. With tight angular acceptance of electrons into the magnet of ~ 1 deg, there is a linear correlation between incident photon energy and electron position recorded on an image plate. Here we present measurements of Bremsstrahlung spectrum from a medical therapy machine, a Scanditronix M22 Microtron. Spectra with energy endpoints from 6 to 20 MeV are directly measured, using a CS with a wide energy range from 0.5 to 20 MeV. We discuss the sensitivity of the device and the effects of converter material and collimation on the accuracy of the reconstructed spectra. Approaches toward improving the sensitivity, including the use of coded apertures, and potential future applications to characterization of spectra are also discussed.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Michelle A. Espy, A. Gehring, A. Belian, T. Haines, J. Hunter, M. James, M. Klasky, J. Mendez, D. Moir, R. Sedillo, R. Shurter, J. Stearns, K. Van Syoc, and P. Volegov "A wide-acceptance Compton spectrometer for spectral characterization of a medical x-ray source", Proc. SPIE 9783, Medical Imaging 2016: Physics of Medical Imaging, 97834V (22 March 2016); https://doi.org/10.1117/12.2216269
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Electrons

Spectroscopy

X-rays

Collimators

Collimation

Magnetism

Coded apertures

RELATED CONTENT

Measuring x-ray spectra of flash radiographic sources
Proceedings of SPIE (August 27 2015)
Recent results from Compton spectrometer experiments
Proceedings of SPIE (September 12 2016)
Intensive x-ray source optimization
Proceedings of SPIE (December 07 2006)

Back to Top